Objective: This phase I IDE study (STAPLE-1) evaluated the primary endpoints of safety (major device-related adverse events at 30 days) and feasibility (successful deployment of all endograft components) of the Aptus Endovascular abdominal aortic aneurysm (AAA) Repair System (Aptus Endosystems, Inc, Sunnyvale, Calif) to treat AAAs.

Methods: A prospective, single arm Federal Drug Administration (FDA) Phase I IDE study was performed. The Aptus endograft is a three-piece modular device with a flexible unsupported main body and two fully supported limbs in a 5.3 mm outer diameter (OD) (16F) delivery system for all iliac limbs and two of three main body sizes. The largest main body (29 mm diameter) is in a 6 mm (18 F OD) delivery system. EndoStaples measuring 4 mm (length) by 3 mm (diameter) designed to provide transmural graft fixation to the adventitia are applied independent of the endograft delivery system. Inclusion criteria included a proximal aortic neck length of 12 mm and iliac landing zone of 10 mm. Secondary endpoints included freedom from endoleaks, rupture, migration, and device integrity.

Results: Twenty-one (21) patients were enrolled at five centers. All patients received the Aptus Endograft and EndoStaples. Ninety-six EndoStaples (range, 2-10; median, 4) were implanted. All patients (n = 21) completed 1-month and 6-month follow-up evaluation and 14 completed 1-year follow-up. Two proximal cuffs and one limb extension were used as adjunctive endograft components at implantation. Three secondary interventions were performed in 2 patients for limb thrombosis. There were no EndoStaple-related adverse events, device integrity failures, migrations, or conversions.

Conclusion: These results of the STAPLE-1 trial document the acute safety and feasibility of the Aptus Endograft and EndoStaples. Early follow-up demonstrates excellent 6-month and 1-year results. A pivotal phase II trial is underway at 25 US centers.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2008.10.064DOI Listing

Publication Analysis

Top Keywords

aptus endograft
12
main body
12
delivery system
12
aptus endovascular
8
repair system
8
phase ide
8
ide study
8
adverse events
8
endograft components
8
endograft endostaples
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!