Y-piece temperature and humidification during mechanical ventilation.

Respir Care

Department of Pulmonary and Critical Care Medicine, Stony Brook University Medical Center, State University of New York at Stony Brook, HSC T 17-040, Stony Brook NY 11794, USA.

Published: April 2009

Background: Practitioners often presume there is adequate humidification in the ventilator circuit if the Y-piece is at a specified temperature, but control of Y-piece temperature may be inadequate to ensure adequate humidification.

Methods: In an in vitro bench model we measured water-vapor delivery with several heated humidification setups and a wide range of minute volume (V (E)) values. The setup included a condenser, hygrometry, and thermometer. First, we calibrated the system with a point-source humidifier and water pump. Then we tested the water-vapor delivery during non-heated-wire humidification and during heated-wire humidification with a temperature gradient of +3 degrees C, 0 degrees C, and -3 degrees C between the humidifier and the Y-piece. We compared the results to 2 recommended humidification values: 100% saturated (absolute humidity 44 mg H(2)O/L) gas at 37 degrees C (saturated/37 degrees C); and 75% saturated (absolute humidity 33 mg H(2)O/L), which is the humidity recommended by the International Organization for Standardization (the ISO standard). In all the experiments the setup was set to provide 35 degrees C at the Y-piece.

Results: Our method for measuring water-vapor delivery closely approximated the amount delivered by a calibrated pump, but slightly underestimated the water-vapor delivery in all the experiments and the whole V (E) range. At all V (E) values, water-vapor delivery during non-heated-wire humidification matched or exceeded saturated/37 degrees C and was significantly greater than that during heated-wire humidification. During heated-wire humidification, water-vapor delivery varied with the temperature gradient and did not reach saturated/37 degrees C at V (E) > 6 L/min. Water-vapor delivery with the negative temperature gradient was below the ISO standard.

Conclusions: Maintaining temperature at one point in the inspiratory circuit (eg, Y-piece), does not ensure adequate water-vapor delivery. Other factors (humidification system, V (E), gradient setting) are critical. At a given temperature, humidification may be significantly higher or lower than expected.

Download full-text PDF

Source

Publication Analysis

Top Keywords

water-vapor delivery
32
y-piece temperature
12
heated-wire humidification
12
temperature gradient
12
saturated/37 degrees
12
humidification
11
temperature humidification
8
circuit y-piece
8
ensure adequate
8
water-vapor
8

Similar Publications

A new fusidic acid-loaded hydrogel film was prepared via the solvent casting technique using alginate and Aloe vera. The hydrogel films were optimized using different ratios of sodium alginate, Aloe vera, and glycerin. The films containing 10% glycerin (w/w of alginate) exhibited the best appearance.

View Article and Find Full Text PDF

Introduction: The increasing prevalence of electronic nicotine delivery systems and alcohol drinking has led to increases in nicotine and alcohol co-use. However, the impact of ENDs on brain activity and binge drinking behavior is not fully understood.

Aims And Methods: We subjected female and male C57BL/6J mice to a voluntary drinking and electronic nicotine vapor exposure paradigm.

View Article and Find Full Text PDF

Investigation of Silver- and Plant Extract-Infused Polymer Systems: Antioxidant Properties and Kinetic Release.

Int J Mol Sci

November 2024

Department of Material Engineering, Faculty of Materials Engineering and Physics, Cracow University of Technology, 31-864 Kraków, Poland.

This study evaluated the impact of silver particles, suspended in flower extract, on the physicochemical characteristics and release dynamics of antioxidant compounds in PVP (polyvinylpyrrolidone)-based hydrogel systems. The hydrogels were synthesized via photopolymerization with fixed amounts of crosslinker (PEGDA) and photoinitiator, while the concentration of the silver-infused extract was systematically varied. Key properties, including the density, porosity, surface roughness, swelling capacity, and water vapor transmission rate (WVTR), were quantitatively analyzed.

View Article and Find Full Text PDF

The increasing prevalence of unintended pregnancies, a persistent issue affecting public health and hindering progress towards the Sustainable Development Goals (SDGs), highlights the critical need for innovative contraceptive approaches. While current methods, including hormonal contraceptives such as levonorgestrel (LNG), offer potential solutions, challenges like limited access and inconsistent use persist. This study introduces a new approach with the development of a three-layer microneedle (TIMN) containing LNG designed to provide extended contraceptive efficacy.

View Article and Find Full Text PDF

Chronic wounds present significant challenges with high morbidity and mortality. A cost-effective dressing that can absorb large exudate volumes, is hemostatic and therapeutically active is of current interest. This study compares two crosslinking approaches on composite scaffolds comprising fish collagen (FCOL), hyaluronic acid (HA) and sodium alginate (SA) by respectively targeting HA and SA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!