Flow uniformity in oxygenators with different outlet port design.

ASAIO J

Consolidated Research Institute for Advanced Science and Medical Care, Waseda University, Okubo, Shinjuku-ku, Tokyo, Japan.

Published: July 2009

This study reports on evaluation of the optimum design of a blood outlet port structure for providing uniform flow by visualizing the blood flow in an extracapillary membrane oxygenator. We tested a cylindrical type extracapillary membrane oxygenator, HPO-20. The HPO-20 has a tangential blood outlet port and is thus referred to as "Tangential HPO-20." We engineered "Vertical HPO-20" with a vertical blood outlet port by modifying the Tangential HPO-20. To visualize the blood-side flow, a total of 120 insulated copper-wire electrodes were placed in the "Tangential" and the Vertical HPO-20s. The test solution flow was visualized by the dimensionless fluid arrival time reaching each electrode. The test solution flow in the Tangential HPO-20 was not uniform, particularly at the outside blood channel. The flow was more uniform in the Vertical HPO-20. The blood flow in an extracapillary membrane oxygenator with a vertical blood outlet port is well distributed so that it produces more uniform blood flow than that with a tangential outlet port because of the small stagnation and reduced channeling.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0b013e31819c6f19DOI Listing

Publication Analysis

Top Keywords

outlet port
24
blood outlet
16
blood flow
12
extracapillary membrane
12
membrane oxygenator
12
flow
9
blood
8
flow extracapillary
8
vertical blood
8
tangential hpo-20
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!