Background: Gold has excellent electrical conductive properties and creates deeper and wider lesions than platinum-iridium during radiofrequency (RF) ablation in vitro. We tested the maximum voltage-guided technique (MVGT) of cavotricuspid isthmus (CTI) ablation using two 8-mm tip catheters containing gold (group G) or platinum-iridium (group PI).

Methods: We enrolled 31 patients who underwent CTI ablation. In group G (n = 15) CTI ablation was performed with a gold-tip ablation catheter, while in group PI (n = 16) a platinum-iridium tip was used. Ablation was guided by CTI potentials with the highest amplitude until achievement of bidirectional block (BIB). If BIB was not achieved after 10 RF applications, RF was delivered via a 3.5-mm irrigated-tip catheter. Success rate, procedure duration, duration of fluoroscopic exposure, and number of RF applications were measured.

Results: BIB was achieved in all patients in group G, while in group PI an irrigated tip was used in four patients (0% vs 25%, P < 0.001). These four patients required a total of 21 additional RF applications (5.25 +/- 2.22). Procedure time (56.4 +/- 12 vs 73.1 +/- 15 minutes P < 0.05) and fluoroscopic explosure (4.9 +/- 2.3 vs 7.1 +/- 3.8 minutes, P < 0.01) were shorter in group G than in group PI. Mean number of RF applications was lower (4.6 +/- 1.9 vs 6.6 +/- 3.1 P < 0.001) and total RF duration shorter (280 +/- 117 vs 480 +/- 310 seconds) in group G than in group PI. No difference was observed in the number of recurrences at a 6 month-follow up (1 in group G vs 1 in group PI).

Conclusion: Using the MVGT of CTI ablation, gold-tip catheters were associated with shorter procedural and fluoroscopic times, and fewer RF applications.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2008.02270.xDOI Listing

Publication Analysis

Top Keywords

cti ablation
16
group group
16
group
12
+/-
9
cavotricuspid isthmus
8
ablation
8
group platinum-iridium
8
bib achieved
8
number applications
8
+/- minutes
8

Similar Publications

Background: Atrial flutter (AFL) is usually effectively treated by cavotricuspid isthmus (CTI) ablation. If AFL recurs despite ablation, there is risk of progression to atrial fibrillation (AF) and clinicians should consider underlying structural heart diseases. This consideration becomes especially critical when right-heart-chambers are dilated.

View Article and Find Full Text PDF

Aims: Cavotricuspid isthmus (CTI) ablation is the current ablation treatment for typical atrial flutter (AFL). However, post-ablation atrial tachyarrhythmias, mostly in the form of atrial fibrillation (AF), are frequently observed after CTI ablation. We aimed to evaluate the effectiveness and safety of concomitant or isolated pulmonary vein isolation (PVI) in patients with typical AFL scheduled for ablation.

View Article and Find Full Text PDF

Background: The aim of cavotricuspid isthmus (CTI)-dependent flutter ablation is the bidirectional conduction block of the CTI. Very-high-power short-duration (vHPSD) radiofrequency ablation aims to minimize conductive heating and increase resistive heating to create shallower but wider lesions in a very short time, while reducing the risk of collateral tissue damage. Experimental studies have shown that it produces effective transmural lesions with an equal or better safety profile compared to conventional parameters.

View Article and Find Full Text PDF

Introduction: The suitability of high-power short-duration (HPSD) cavo-tricuspid isthmus ablation (CTI-Abl) for electrophysiology (EP) trainees, as well as the underlying mechanisms of its efficacy, remain unknown. The aim of this study was to clarify the efficacy and safety of HPSD CTI-Abl performed by EP trainees and assess lesion characteristics between HPSD and moderate-power long duration (MPLD) ablations.

Methods: Study 1: CTI-Abl was performed by first- to fourth-year EP trainees in consecutive 113 patients (67 ± 11 years, 27.

View Article and Find Full Text PDF

Background: Cardiac magnetic resonance (CMR)-guided catheter ablation of the cavotricuspid isthmus (CTI) has been proven feasible, but determinants of local electrogram (EGM) voltage drops during radiofrequency (RF) applications are unknown.

Objective: The purpose of this study was to investigate local atrial bipolar EGM voltage drops and the association with delivered RF energy and anatomical information derived from peri-procedural CMR imaging.

Methods: In consecutive patients undergoing CMR-guided CTI ablation procedures, relative EGM voltage drops for RF applications ≥20 seconds were calculated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!