Background: In automated erythrocytapheresis procedures, achieving the desired-end hematocrit (Hct) requires that the COBE Spectra (CaridianBCT) machine be programmed with the mean Hct of the replacement red blood cell (RBC) units. To determine unit Hct, data derived from quality control (QC) Hct data were utilized. However, if a unit volume is outside the quality control (QC) volume parameters, the unit is accessed to measure its Hct. In this study, 21 percent of all RBC units need to be accessed to determine the Hct, which affects 47.5 percent of patient's erythrocytapheresis procedures. Spiking the unit compromises its integrity and hastens the expiration time of the unit. Nurses must wait until the patient arrives to check units outside QC parameters, thereby delaying the start time of procedures. Even if sampled units are kept refrigerated, they cannot be returned to the blood bank inventory once spiked. The goal of this study was to determine if accurate Hct levels from RBC units could be obtained from a unique segment.

Study Design And Methods: To determine the centrifuged Hct, samples were prepared from the RBC units and compared to that of the unique segment.

Results: The Hct of the unique segment exceeded that of the RBC unit by a small (1.2% in AS-1 units, 0.92% in AS-3 units), but statistically significant amount.

Conclusion: The Hct from unique RBC segments closely approximates that of the original RBC unit. Unique segments can be made that will maintain the integrity and shelf life of RBC units.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1537-2995.2009.02099.xDOI Listing

Publication Analysis

Top Keywords

rbc units
20
units
10
hct
10
red blood
8
blood cell
8
unit
8
erythrocytapheresis procedures
8
rbc
8
hct data
8
quality control
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!