Validation study of a portable monitoring device for identifying OSA in a symptomatic patient population.

Respirology

Respiratory Medicine Division, Department of Medicine and Therapeutics, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong, China.

Published: March 2009

Background And Objective: Obstructive sleep apnoea syndrome (OSAS) is a common disorder associated with early atherosclerosis, diabetes mellitus, ischaemic heart disease and cerebrovascular disease. The gold standard for confirming OSAS is based on an attended overnight polysomnography (PSG) in a sleep laboratory; however lack of health-care resources creates long waiting times for patient access to this diagnostic test. This study evaluated the ability of a portable sleep-monitoring device to identify patients in Hong Kong with suspected OSAS.

Methods: Patients with symptoms of OSAS were invited to use the ARES (apnoea risk evaluation system) concurrently with an attended inpatient PSG. Several sets of AHI were generated by the ARES provider based on different oxygen desaturation criteria and surrogate parameters of arousal. The results were compared against PSG to determine the optimal sensitivity and specificity.

Results: There were 141 patients who completed the study successfully. Results of AHI from the ARES study were presented in the order of different scoring criteria--4% oxygen desaturation alone, obstructive events with 3% oxygen desaturation and obstructive events with 1% desaturation plus surrogate arousal criteria. The sensitivity was 0.84 (95% confidence interval (CI): 0.77-0.90), 0.89 (95% CI: 0.89-0.94) and 0.97 (95% CI: 0.94-0.99), respectively. The specificity was 1, 1 and 0.63 (95% CI: 0.55-0.71), respectively. The receiver operating curve had an area of 0.96, 0.97 and 0.98, respectively. The kappa coefficient varied from 0.24 to 0.55 for agreement of severity between PSG and ARES. The likelihood ratio positive and the likelihood ratio negative were 2.61, infinity, infinity and 0.16, 0.11, 0.05, respectively, in the order of oxygen desaturation described earlier.

Conclusions: The ARES device has reasonable sensitivity and specificity for diagnosing severe OSAS in symptomatic Chinese patients. There is moderate agreement between ARES and PSG in the diagnosis of severe disease, but less agreement in patients with mild/moderate disease.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1440-1843.2008.01439.xDOI Listing

Publication Analysis

Top Keywords

oxygen desaturation
16
desaturation obstructive
8
obstructive events
8
likelihood ratio
8
ares
6
psg
5
patients
5
desaturation
5
validation study
4
study portable
4

Similar Publications

Massive endobronchial hemorrhage leading to Cardiac arrest during EBUS-TBNA: a case of successful resuscitation.

BMC Pulm Med

January 2025

Department of Pulmonary and Critical Care Medicine, School of Medicine, Zhongshan Hospital of Xiamen University, Xiamen University, Xiamen, Fujian, China.

Introduction: Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is commonly used for diagnosing mediastinal lymphadenopathy. Despite a low complication rate, severe hemorrhage can occur which is reported in this literature, particularly in hypervascular conditions like Castleman disease.

Methods: A 54-year-old male with idiopathic multicentric Castleman disease underwent EBUS-TBNA for mediastinal lymph node sampling.

View Article and Find Full Text PDF

: Obstructive sleep apnea is a sleep-related breathing disorder associated with craniofacial morphology and dental arches. The aim of this study was to evaluate the correlation between obstructive sleep apnea and the morphometry of dental arches and upper airways. : Forty patients were enrolled in the study, and the polysomnographic parameters evaluated were the apnea hypopnea index (AHI) and the oxygen desaturation index (ODI).

View Article and Find Full Text PDF

Mechanism of Transcription Factor ChbZIP1 Enhanced Alkaline Stress Tolerance in .

Int J Mol Sci

January 2025

State Key Laboratory of Microbial Metabolism, School of Life Sciences and Biotechnology, Shanghai Jiao Tong University, Shanghai 200240, China.

Alkaline environments such as alkaline lands, lakes, and industrial wastewater are not conducive to the growth of plants and microorganisms due to high pH and salinity. ChbZIP1 is a bZIP family transcription factor isolated from an alkaliphilic microalgae ( sp. BLD).

View Article and Find Full Text PDF

Background: Rapid sequence induction intubation (RSII) is commonly used in emergency surgeries for patients at high risk of aspiration. However, these patients are more susceptible to hypoxemia during the RSII process. High-flow nasal cannula (HFNC) oxygen therapy has emerged as a potential alternative to traditional face mask (FM) ventilation pre- and apneic oxygenation.

View Article and Find Full Text PDF

Cardiopulmonary bypass (CPB) in cardiac surgery is associated with a high risk of postoperative neurological complications. Perioperative use of vasopressors is common to counteract arterial hypotension in this setting. However, use of α-agonist vasopressors has been associated with cerebral desaturations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!