A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A novel device for protecting rectum during prostate cancer irradiation: in vivo data on a large mammal model. | LitMetric

Purpose: Hypofractionation schemes and associated higher rectal doses have evoked the need for improved protection of the rectum during prostate cancer irradiation.

Materials And Methods: An implantable, biodegradable, inflatable, preshaped triangular balloon of commercially used poly(L-lactide-co-epsilon-caprolactone) co-polymer material was developed to provide separation between prostate and rectum. Biocompatibility and degradability of the balloon implanted subcutaneously or perineally, and in the context of transperineal implantation and local irradiation were evaluated in several in vivo studies.

Results: The device was found to be biocompatible in subcutaneously implanted rabbits up to 42 days, in a transperineally implanted dog up to 12 months and in 8 transperineally implanted pigs up to 6 months. Upon inflation in situ the balloon separated the tissues, remained inflated for several months and subsequently biodegraded. No systemic or local toxicity was noted, as shown by histopathology. Device insertion into the perineal area using a dedicated introductory kit was convenient and feasible. Three-month followup in irradiated pigs that received 15 Gy in 3 fractions 1 week apart showed a stable balloon position with no local or systemic side effects.

Conclusions: This novel device was safe and effective for its intended use of separating tissues for a desired duration. A clinical study will commence to evaluate the safety and efficacy of this device during irradiation in patients with prostate cancer.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.juro.2008.11.010DOI Listing

Publication Analysis

Top Keywords

prostate cancer
12
novel device
8
rectum prostate
8
transperineally implanted
8
device protecting
4
protecting rectum
4
prostate
4
cancer irradiation
4
irradiation vivo
4
vivo data
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!