The application of the thermal desorption (TD) method coupled with gas chromatography-mass spectrometry (GC-MS) to the analysis of aerosol organics has been the focus of many studies in recent years. This technique overcomes the main drawbacks of the solvent extraction approach such as the use of large amounts of toxic organic solvents and long and laborious extraction processes. In this work, the application of an automatic TD-GC-MS instrument for the determination of particle-bound polycyclic aromatic hydrocarbons (PAHs) is evaluated. This device offers the advantage of allowing the analysis of either gaseous or particulate organics without any modification. Once the thermal desorption conditions for PAH extraction were optimised, the method was verified on NIST standard reference material (SRM) 1649a urban dust, showing good linearity, reproducibility and accuracy for all target PAHs. The method has been applied to PM10 and PM2.5 samples collected on quartz fibre filters with low volume samplers, demonstrating its capability to quantify PAHs when only a small amount of sample is available.