Aims: Currently available devices for transcatheter closure of patent foramen ovale (PFO) which rely on a permanent implant have limitations, including late complications. The study objective was to evaluate the safety, feasibility, and effectiveness of the PFx Closure System, the first transcatheter technique for PFO closure without an implantable device.

Methods And Results: A prospective study of 144 patients was conducted at nine clinical sites from October 2005 through August 2007. All patients had a history of cryptogenic stroke, transient ischemic attack, migraines, or decompression illness. The mean balloon stretched diameter of the PFO was 7.9 +/- 2.5 mm. Technical success (successful application of radiofrequency energy) was achieved in 130 patients. One patient required a transfusion as a result of blood loss during the procedure. There were no other major procedural complications. There were no recurrent strokes, deaths, conduction abnormalities, or perforations following the procedure. At a mean follow-up of 6 months, successful closure was achieved in 79 patients (55%). In PFOs with balloon sized or stretched diameters less than 8 mm, the closure rate was 72% (53/74).

Conclusion: This study demonstrates that transcatheter closure of a PFO without a permanent implant is technically feasible and safe. Further technique and device modifications are required to achieve higher closure rates.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.21809DOI Listing

Publication Analysis

Top Keywords

transcatheter closure
12
closure patent
8
patent foramen
8
foramen ovale
8
144 patients
8
permanent implant
8
closure
7
patients
5
transcatheter
4
ovale radiofrequency
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!