Clinically deployable Kinesia technology for automated tremor assessment.

Mov Disord

Division of Movement Disorders, Cleveland Medical Devices Inc., Cleveland, Ohio 44103, USA.

Published: April 2009

The objective was to design, build, and assess Kinesia, a wireless system for automated assessment of Parkinson's disease (PD) tremor. The current standard in evaluating PD is the Unified Parkinson's Disease Rating Scale (UPDRS), a qualitative ranking system typically completed during an office visit. Kinesia integrates accelerometers and gyroscopes in a compact patient-worn unit to capture kinematic movement disorder features. Objectively quantifying PD manifestations with increased time resolution should aid in evaluating efficacy of treatment protocols and improve patient management. In this study, PD subjects performed the tremor subset of the UPDRS motor section while wearing Kinesia. Quantitative kinematic features were processed and highly correlated to clinician scores for rest tremor (r(2) = 0.89), postural tremor (r(2) = 0.90), and kinetic tremor (r(2) = 0.69). The quantitative features were used to develop a mathematical model that predicted tremor severity scores for new data with low errors. Finally, PD subjects indicated high clinical acceptance.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mds.22445DOI Listing

Publication Analysis

Top Keywords

parkinson's disease
8
tremor
7
clinically deployable
4
kinesia
4
deployable kinesia
4
kinesia technology
4
technology automated
4
automated tremor
4
tremor assessment
4
assessment objective
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!