  • Miniaturization of analytical separation methods enhances efficiency by providing shorter run times, higher resolution, and better recovery of samples.
  • A new approach is presented to quantify reversible adsorption without the need for a specific equation, using two experiments to achieve comparable variance under different conditions.
  • The findings reveal significant protein loss (40-50%) due to hydrophobic interactions from minimal coating modifications, with better reproducibility observed in certain capillary setups.

Article Abstract

Miniaturization of analytical separation methods offers several advantages, including short run times, high resolution, and high recovery of the sample constituents. To optimize these parameters, the reversible adsorption (to minimize loss in resolution), as well as the irreversible adsorption (to minimize loss of analytes) must be quantified.However, no useful equation is available for the calculation of the variance of reversible adsorption. Therefore,we have taken another approach to quantify the reversible interaction. The method is unique and important since no equation for calculation of this variance is required. Instead, two experiments are required, which are run under such conditions that the variance of a certain parameter has the same numerical value in the two experiments (one with and without EOF), except for the variance of reversible adsorption. The approach is universal in the sense that it can be used for many different mathematical concepts and be modified to also cover certain functions other than a sum of parameters.We have also introduced a simple expression for their reversible adsorption, which shows that the hydrophobic interaction from only two methyl groups in the coating gives rise to as much as 40-50% loss of protein, and the width of the zones in the capillary with this coating was 8-15% larger compared to the zone width in the polyacrylamide-coated capillaries. The reproducibility in migration time, peak area, and peak width in two consecutive runs in capillaries with two methyl groups in the coating was very low, but in EOF-free polyacrylamide-coated capillaries extremely high, indicating that the reversible and irreversible adsorption of proteins to this coating is negligible. The scanning detector, frequently used in free zone electrophoresis in the 1960s-1970s, gives true separation parameters and is, therefore, much preferable to the stationary detector used in most CE experiments, because this detector gives apparent separation parameters.

Download full-text PDF

Source
http://dx.doi.org/10.1021/ac8010457DOI Listing

Publication Analysis

Top Keywords

reversible adsorption
20
variance reversible
12
adsorption minimize
8
minimize loss
8
irreversible adsorption
8
equation calculation
8
calculation variance
8
methyl groups
8
groups coating
8
polyacrylamide-coated capillaries
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!