Introduction And Objective: The aim of this study was to evaluate the ability of root canal cleaning of the RaCe rotary instrumentation system (FKG Dentaire - La-Cheaux-de Fonds - Switzerland), compared to manual filing with Stainless Steel K-files (Maillefer Instruments - Ballaigues - Switzerland).

Material And Method: Twenty extracted human teeth (maxillary molars) were selected and their pulp tissue was removed after coronal access. The root canals were filled with a dye (India ink) and allowed to dry for 48 hours, followed by establishment of the working length of the mesiobuccal root; then, half of the specimens were instrumented by the modified Oregon technique and the other half were instrumented by crown-down sequence of the RaCe system. After preparation, the teeth were longitudinally sectioned and evaluated according to the amount of remaining dye.

Results: Data obtained were registered as numerical scores, and the arithmetic means were compared between groups using the Mann Whitney test. Both techniques were unable to completely clean the interior of the root canals, with a better performance of the manual technique only at the middle third.

Conclusion: It could be concluded that the RaCe system was able to provide satisfactory cleaning, similar to that obtained by the manual instrumentation technique.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4327163PMC
http://dx.doi.org/10.1590/s1678-77572006000100002DOI Listing

Publication Analysis

Top Keywords

race system
12
root canal
8
root canals
8
root
5
comparison cleaning
4
cleaning efficacy
4
efficacy fkg
4
race
4
fkg race
4
system
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!