Background/objective: Little is known about the long-term effects of chronic exposure to ionizing radiation. Studies have shown that spine surgeons may be exposed to significantly more radiation than that observed in surgery on the appendicular skeleton. Computer-assisted image guidance systems have been shown in preliminary studies to enable accurate instrumentation of the spine. Computer-assisted image guidance systems may have significant application to the surgical management of spinal trauma and deformity. The objective of this study was to compare C-arm fluoroscopy and computer-assisted image guidance in terms of radiation exposure to the operative surgeon when placing pedicle screw-rod constructs in cadaver specimens.

Methods: Twelve single-level (2 contiguous vertebral bodies) lumbar pedicle screw-rod constructs (48 screws) in 4 fresh cadavers were placed using standard C-arm fluoroscopy and computer-assisted image guidance (Stealth Station with Iso-C(3D)). Pedicle screw-rod constructs were placed at L1-L2, L3-L4, and L5-S1 in 4 fresh cadaver specimens. Imaging was alternated between C-arm fluoroscopy and computer-assisted image guidance with StealthStation Iso-C(3D). Radiation exposure was measured using ring and badge dosimeters to monitor the thyroid, torso, and index finger. Postprocedure CT scans were obtained to judge accuracy of screw placement.

Results: Mean radiation exposure to the torso was 4.33 +/- 2.66 mRem for procedures performed with standard fluoroscopy and 0.33 +/- 0.82 mRem for procedures performed with computer-assisted image guidance. This difference was statistically significant (P= 0.012). Radiation exposure to the index finger and thyroid was negligible for all procedures. The accuracy of screw placement was similar for both techniques.

Conclusions: Computer-assisted image guidance systems allow for the safe and accurate placement of pedicle screw-rod constructs with a significant reduction in exposure to ionizing radiation to the torso of the operating surgeon.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2607125PMC
http://dx.doi.org/10.1080/10790268.2008.11753648DOI Listing

Publication Analysis

Top Keywords

computer-assisted image
32
image guidance
32
radiation exposure
20
fluoroscopy computer-assisted
16
pedicle screw-rod
16
screw-rod constructs
16
guidance systems
12
c-arm fluoroscopy
12
lumbar pedicle
8
screw placement
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!