A two year old, female spayed border collie presented three weeks after permanent pacemaker implantation for weakness, lethargy and collapse. Electrocardiogram documented complete (3(rd) degree) atrioventricular (AV) block, the absence of pacing and a ventricular escape rhythm. Thoracic radiographs revealed retraction and dislodgement of the passive fixation lead from the right ventricular apex. During a procedure to reposition the lead, it was noted that lead retraction had resulted from rotation of the generator with twisting and entanglement of the lead on itself and around the generator, a complication recognized in human patients as "Twiddlers syndrome". The pacemaker lead was removed, a new lead securely placed at the right ventricular apex, and the generator was secured into a revised subcutaneous pocket. Pacing was re-established and the Twiddler's syndrome has not reoccurred.

Download full-text PDF

Source
http://dx.doi.org/10.1016/S1760-2734(06)70063-7DOI Listing

Publication Analysis

Top Keywords

twiddler's syndrome
8
ventricular apex
8
lead
6
secure pacemaker
4
pacemaker fixation
4
fixation critical
4
critical prevention
4
prevention twiddler's
4
syndrome year
4
year female
4

Similar Publications

: Deep brain stimulation (DBS) is an effective treatment for movement disorders, but its long-term efficacy may be undermined by hardware complications such as lead fractures. These complications increase healthcare costs and necessitate surgical revisions. The frequency, timing, and clinical factors associated with lead fractures remain poorly understood.

View Article and Find Full Text PDF
Article Synopsis
  • Twiddler's syndrome is a rare condition associated with cardiac pacemakers, characterized by the twisting and retraction of the device lead within the implantation site.
  • A 54-year-old woman was admitted to the emergency department due to chest pain and involuntary muscle contractions in her left side.
  • A cardiology specialist confirmed the diagnosis of Twiddler's syndrome after evaluating the patient's symptoms.
View Article and Find Full Text PDF

Twiddler syndrome is a rare entity in which patient manipulation causes lead dislocation with lead retraction. Reel syndrome, on the other hand, is observed after external manipulation, intentional or unintentional, in which the leads move along their transverse axis and is considered a variant of Twiddler syndrome. We present the clinical case of a 91-year-old female who, after pacemaker implantation, presented with dizziness and chest discomfort following constant manipulation of the pacemaker, resulting in complete retraction of the lead tip into the pouch, which is unusual in the literature to date.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!