A flow injection on-line sorption system was developed for the separation and preconcentration of traces of Ag, Cd, Co, Ni, Pb, U and Y from natural water samples with subsequent detection by ICP TOF MS. Simultaneous preconcentration of the analytes was achieved by complexation with the chelating reagent 1-phenyl-3-methyl-4-benzoylpyrazol-5-one immobilized on the inner walls of a (200cm x 0.5mm) PTFE knotted reactor. The analytes were eluted and transported to an axial ICP TOF MS system with 1% (v/v) HNO(3) containing 0.3microg l(-1) of Rh as an internal standard using ultrasonic nebulization. The detection limits (3sigma) varied from 0.3ng l(-1) for Y to 15.2ngl(-1) for Ni and the precision (R.S.D.) was better than 4%. Using a loading time of 90s and a sample flow rate of 4.5ml min(-1), enhancement factors of 3-14 were obtained for the different analytes in comparison with their direct determination by ICP TOF MS with ultrasonic nebulization without preconcentration. The accuracy of the method was demonstrated by analysis of water based certified reference materials.