Objective: To analyze the reasons of cardiac permanent pacemaker replacement and the strategies to prevent relevant complications.
Methods: The clinical data of 57 patients, 38 males and 19 females, aged 74 +/- 8 (56-94), 31 with sick sinus syndrome, 26 with II or III degree atrioventricular block, who underwent 63 times of cardiac permanent pacemaker replacement, including 13 times of lead replacement, were analyzed.
Results: The reason of replacement included battery normal exhaustion for 57 case-times, battery exhaustion before the defined schedule for 2 case-times, lead electrodes incompletely fractured for 2 case-times, and infection of pacemaker pocket for 2 case-times. There were 9 cases of placement-related complications: pacemaker pocket hematoma (n=4), lead dislocation (n=3), and abandoned leads falling into the right ventricle (n=2). The average lifetime of old pacemakers was 9. 25 years (2 -15 years). The pacing threshold of ventricular leads after pacemaker replacement was (0.77 +/- 0. 40)V, significantly higher than the initial pacing threshold [(0.60 +/- 0.21)V, P < 0.01]. There were no significant differences in the lead impedance and R wave amplitude between the pacemaker replacement and initial implantation [(854 +/- 136)omega vs. (828 +/- 176)omega, and (12 +/- 4)mV vs. (12 +/- 4)mV, both P > 0.05].
Conclusion: The main reason of pacemaker replacement is battery exhaustion. Most implanted ventricular leads still can be used. A rare serious complication of cardiac pacemaker replacement operation is abandoned lead falling into the right ventricle, and correct disposing of initial leads help avoid this complication .
Download full-text PDF |
Source |
---|
Catheter Cardiovasc Interv
January 2025
Division of Cardiovascular Diseases, Bridgeport Hospital, Yale New Haven Health, Bridgeport, Connecticut, USA.
Background: The co-existence of severe aortic stenosis (AS) and hypertrophic cardiomyopathy (HCM) is not uncommon. Surgical intervention is the gold standard management. Patients with high surgical risk might undergo transcatheter aortic valve replacement (TAVR).
View Article and Find Full Text PDFIndian Pacing Electrophysiol J
December 2024
Intermountain Heart Institute - Utah Valley Hospital, Salt Lake City, Utah, USA.
Introduction: The advancement of medical technology has introduced leadless pacemakers (LPMs) as a significant innovation in cardiac pacing, offering potential advantages over traditional ventricular transvenous pacemakers. This report explores the application of LPMs in two patients with complex valvular histories, particularly those with mechanical tricuspid valves.
Case Reports: The first case involves a 60-year-old male with a history of rheumatic heart disease and triple valve replacement who developed a high-grade AV block.
J Cardiothorac Surg
December 2024
Department of Cardiology, Thomas Jefferson University, Philadelphia, PA, 19107, USA.
Background: Right ventricular (RV) function assessment by echocardiography can be challenging due to its complex morphology. Also, increasing use of sedation rather than general anesthesia for transfemoral approach transcatheter aortic valve replacement (TAVR) reduces the need for intraoperative transesophageal echocardiography (TEE). Recent clinical studies have demonstrated the importance of 3-dimensional (3D) echocardiography and a longitudinal strain for RV function assessment.
View Article and Find Full Text PDFBMJ Open
December 2024
Graduate Institute of Sport, Leisure and Hospitality Management, National Taiwan Normal University, Taipei, Taiwan.
Objective: Phase angle (PhA) is a prognostic factor for predicting and monitoring geriatric syndromes. However, multiple factors associated with increased PhA values as an outcome remain unclear in the older population. This study aimed to examine the association of socio-demographic, anthropometric and behavioural factors with PhA among older Taiwanese adults.
View Article and Find Full Text PDFJ Cardiovasc Dev Dis
December 2024
Faculty of Medicine & Health Sciences, University of Antwerp, 2610 Antwerp, Belgium.
The need for a permanent pacemaker (PPM) implantation after surgical aortic valve implantation (SAVR) is a recognized postoperative complication, with potentially long-term reduced survival. From 1987 to 2017, 2500 consecutive patients underwent SAVR with a biological valve with or without concomitant procedures such as CABG or mitral valve repair. Mechanical valves or valves in another position were excluded.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!