Absolute calibration of microfission chamber in JT-60U.

Rev Sci Instrum

Japan Atomic Energy Agency, Mukouyama 801-1, Naka, Ibaraki 311-0193, Japan.

Published: October 2008

We have conducted calibrations of a microfission chamber, which was installed between the vacuum vessel and the toroidal field coils, by both a Cf-252 neutron source and real deuterium plasmas in JT-60U. The detector employs both pulse counting and Campbell (mean square voltage) modes in the electronics to cover a wide dynamic range of the neutron source strength. The pulse counting and Campbell modes were calibrated by Cf-252 and deuterium plasmas, respectively. Point efficiencies, counts per neutron from a point at a single angle, were measured for 27 locations of the neutron point source in toroidal scan. The efficiencies were influenced by the various components such as the vacuum vessel, port, and graphite tiles. The point efficiencies can be integrated and averaged with angle to provide toroidal line efficiencies. The line efficiencies of the microfission chamber and the nearest neutron monitor of the U-235 fission chamber were 5.38x10(-9) and 1.77x10(-8), respectively. Then the calibration for the Campbell mode was also performed by using a real deuterium plasma. The detection efficiency in the Campbell mode was about three-tenths of that of the nearest neutron monitor, which is consistent with the calibration result obtained by using a Cf-252 neutron source for pulse counting mode.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.2968709DOI Listing

Publication Analysis

Top Keywords

microfission chamber
12
neutron source
12
pulse counting
12
vacuum vessel
8
cf-252 neutron
8
real deuterium
8
deuterium plasmas
8
counting campbell
8
point efficiencies
8
neutron point
8

Similar Publications

Development of in-vessel components of the microfission chamber for ITER.

Rev Sci Instrum

October 2010

Japan Atomic Energy Agency, 801-1, Mukoyama, Naka 311-0193, Japan.

Microfission chambers (MFCs) will measure the total neutron source strength in ITER. The MFCs will be installed behind blanket modules in the vacuum vessel (VV). Triaxial mineral insulated (MI) cables will carry signals from the MFCs.

View Article and Find Full Text PDF

Design of microfission chamber for ITER operations.

Rev Sci Instrum

October 2008

Fusion Research and Development Directorate, Japan Atomic Energy Agency, Ibaraki 311-0193, Japan.

Microfission chambers (MFCs) are one of the most important diagnostics for measuring neutron source strength in ITER. Using MFCs for high-power operations (fusion power of 100 kW-1 GW) and for low-power operations (<100 kW) in combination is one way to fulfill the target measurement requirements of ITER. The MFCs for high-power operations will be installed behind blanket modules in both the upper and lower outboard regions of the vacuum vessel so as to be insensitive to changes in the position of the plasma.

View Article and Find Full Text PDF

Absolute calibration of microfission chamber in JT-60U.

Rev Sci Instrum

October 2008

Japan Atomic Energy Agency, Mukouyama 801-1, Naka, Ibaraki 311-0193, Japan.

We have conducted calibrations of a microfission chamber, which was installed between the vacuum vessel and the toroidal field coils, by both a Cf-252 neutron source and real deuterium plasmas in JT-60U. The detector employs both pulse counting and Campbell (mean square voltage) modes in the electronics to cover a wide dynamic range of the neutron source strength. The pulse counting and Campbell modes were calibrated by Cf-252 and deuterium plasmas, respectively.