Objectives: To evaluate the efficacy of the computer-aided detection (CAD) system and thin-slab maximum intensity projection (MIP) technique in the detection of pulmonary nodules at multidetector computed tomography (CT) in patients who underwent metastatectomy.

Materials And Methods: This retrospective study was approved by the institutional review board and patients' informed consent was waived. Forty-nine consecutive patients who underwent pulmonary metastatectomy were enrolled. Four chest radiologists analyzed preoperative 1-mm section CT images and recorded the locus of each nodule candidate. Afterward, they reevaluated the images once using CAD software and once with thin-slab MIP given the results of 1-mm section CT alone. The reference standard for nodule presence was established by a consensus panel and pathologic records for malignant nodules.

Results: A total of 514 nodules were identified by a consensus panel. Of 212 nodules surgically removed, 121 nodules were malignant. The sensitivity of each observer in detecting malignant nodules with thin-section CT scans alone was 91%, 88%, 87%, and 86% for observers A- to D, respectively. With CAD, sensitivity increased significantly to 95%, 95%, 94%, and 95% (P< 0.05 for observer B-D), and using MIP increased to 94%, 96%, 91%, and 92% (P < 0.05 for observer B-D), respectively. There were no significant differences in sensitivity between CAD and MIP for the detection of malignant nodules. The average number of false-positive findings per patient was 0.8 with thin-section CT alone, 1.1 with CAD, and 1.4 with MIP.

Conclusions: In candidates for metastatectomy, reading with the aid of either CAD or MIP significantly improved the detection of malignant nodules compared with using thin-section CT alone.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0b013e318190fcfcDOI Listing

Publication Analysis

Top Keywords

malignant nodules
12
efficacy computer-aided
8
computer-aided detection
8
system thin-slab
8
thin-slab maximum
8
maximum intensity
8
intensity projection
8
technique detection
8
detection pulmonary
8
nodules
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!