Background: We compared the analytical and clinical performance of 3 porcine thyroid receptor antibody (TRAb) methods (1 second- and 2 new third-generation systems) with the conventional TRAb assay based on the human recombinant TSH receptor (hTRAK).

Patients And Methods: We obtained sera from 86 patients with untreated Graves disease (GD) and 71 healthy controls. We measured TRAb concentrations by radioreceptor assay using the hTRAK (Brahms) or the porcine TSH receptor (pRRA) from Beckman-Coulter, by electrochemiluminescence immunoassay (ECLIA) with the Elecsys/Cobas (Roche), and by ELISA using the Medizym TRAb clone (Medipan).

Results: Between-run assay imprecision was < or =10% and < or =7.6% for hTRAK and ECLIA, but reached 14% and 14.9% for ELISA and pRRA, respectively. Maximal specificity and sensitivity close to 100% were obtained for hTRAK, ECLIA, and ELISA. pRRA failed to detect positive TRAbs in 5 GD patients. Although calibrated against the same reference standard 90/672, the assays displayed a high intermethod variability. The results were significantly higher by ECLIA and lower by ELISA and pRRA compared with hTRAK. Patients with ophthalmopathy had higher TRAb results by ELISA and pRRA than those without eye disease.

Conclusions: Second- and third-generation TRAb assays had similar diagnostic sensitivities in the diagnostic evaluation of GD. Despite the use of the same reference standard for calibration, high intermethod variability in TRAb assay results was seen in untreated GD patients. Assay harmonization is necessary for correct interpretation in the follow-up of Graves ophthalmopathy.

Download full-text PDF

Source
http://dx.doi.org/10.1373/clinchem.2008.115162DOI Listing

Publication Analysis

Top Keywords

elisa prra
16
intermethod variability
12
graves disease
8
follow-up graves
8
graves ophthalmopathy
8
second- third-generation
8
trab assay
8
tsh receptor
8
htrak eclia
8
reference standard
8

Similar Publications

The objective of this study was to assess whether an antibody ELISA applied to bulk tank milk (BTM) could be used to accurately estimate within-herd prevalence of digital dermatitis (DD). The ELISA was designed for the detection of antibodies against Treponema phagedenis-like strain V1 (PrrA antigen). The hind feet of all lactating cows from 40 commercial French dairy herds with a history of DD were scored by an observer in the milking parlor, using the 4 M-stage system.

View Article and Find Full Text PDF

Digital dermatitis (DD) is a contagious claw disease causing lameness in cattle, affecting both animal welfare and economics. In this study, shotgun phage display was used to identify immunogenic proteins in a strain (V1) of the Treponema phylotype closely related to Treponema phagedenis, indicated as a key agent in the pathogenesis of DD. A genomic phage library was constructed and selected against antibodies from a rabbit immunized with live strain V1 bacteria.

View Article and Find Full Text PDF

Background: We compared the analytical and clinical performance of 3 porcine thyroid receptor antibody (TRAb) methods (1 second- and 2 new third-generation systems) with the conventional TRAb assay based on the human recombinant TSH receptor (hTRAK).

Patients And Methods: We obtained sera from 86 patients with untreated Graves disease (GD) and 71 healthy controls. We measured TRAb concentrations by radioreceptor assay using the hTRAK (Brahms) or the porcine TSH receptor (pRRA) from Beckman-Coulter, by electrochemiluminescence immunoassay (ECLIA) with the Elecsys/Cobas (Roche), and by ELISA using the Medizym TRAb clone (Medipan).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!