A single column, rapid quality control procedure for 6-[18F]fluoro-L-dopa and 6-[18F]fluorodopamine PET imaging agents.

Int J Rad Appl Instrum B

Nuclear Medicine Department, Warren G. Magnuson Clinical Center, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD 20892.

Published: June 1991

6-[18F]Fluoro-L-dopa and 6-[18F]fluorodopamine are promising PET imaging agents for visualizing cerebral dopaminergic centers and cardiac sympathetic innervation and function. Administration to humans requires a means to determine the purity before injection. We describe such a method using HPLC with u.v. and radioactivity detection and a single high-speed C-18 column with gradient elution. The procedure can resolve within 10 min these fluorinated catechols, their isomers, and dihydroxyphenylalanine. The chemical and radiochemical purity, and specific activity, can be determined before injection.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0883-2897(91)90081-uDOI Listing