Background: We evaluated the analytical characteristics of electrochemiluminescence (ECLIA) immunoassay for NT-proBNP (proBNP II, Roche Diagnostics, Germany) and compared its analytical performance to that of the previous polyclonal method.

Methods: We measured NT-proBNP in EDTA plasma samples of 177 consecutive cardiac patients (69 females and 108 males; mean age 62.2+/-16.4 years, range 13 to 96 years) with monoclonal and polyclonal ECLIA methods following manufacturer's instructions using an Elecsys 2010 analyzer.

Results: Monoclonal ECLIA method for NT-proBNP assay showed an imprecision (CV%) lower than 3% at the cut-off value (i.e., 150 ng/L). No significant interference was found in plasma samples containing high levels of hemoglobin, triglycerides or bilirubin. EDTA plasma showed slightly, but significantly lower NT-proBNP values than serum (on average -6.3%) and lithium-heparinized plasma (on average -3.9%) samples. Finally, a very close linear regression was found between the NT-proBNP values found by either monoclonal or polyclonal ECLIA method (monoclonal ECLIA=-72.17+1.04 polyclonal ECLIA, n=177, R=0.993).

Conclusions: The monoclonal ECLIA method showed very similar analytical characteristics with slightly lower NT-proBNP results (on average -2.5%) than the polyclonal ECLIA method. The difference between monoclonal and polyclonal methods seems to be too slight to change the reference range and decisional values for NT-proBNP assay, as previously assessed by the polyclonal ECLIA method.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2008.10.011DOI Listing

Publication Analysis

Top Keywords

polyclonal eclia
20
eclia method
20
monoclonal polyclonal
12
polyclonal
8
nt-probnp
8
analytical characteristics
8
eclia
8
edta plasma
8
plasma samples
8
monoclonal eclia
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!