All four currently FDA-approved drug-eluting stents (DESs) contain a durable polymeric coating which can negatively impact vascular healing processes and eventually lead to adverse cardiac events. Aim of this study was the pre-clinical assessment of two novel rapamycin-eluting stent (RES) coating technologies that abstain from use of a durable polymer. Two distinctive RES coating technologies were evaluated in vitro and in the porcine coronary artery stent model. The R-poly(S) stent platform elutes rapamycin from a biodegradable polymer that is top coated with the resin shellac to minimize the amount of polymer. The R-pro(S) stent platform allows dual drug release of rapamycin and probucol, blended by shellac. HPLC-based determination of pharmacokinetics indicated drug release for more than 28 days. At 30 days, neointimal formation was found to be significantly decreased for both DESs compared to bare-metal stents. Assessment of vascular healing revealed absence of increased inflammation in both DESs, which is commonly observed in DES with non-erodible polymeric coating. In conclusion, the pre-clinical assessment of RESs with resin-based or dual drug coating indicated an adequate efficacy profile as well as a beneficial effect for vascular healing processes. These results encourage the transfer of these technologies to clinical evaluation.