Objective: To evaluate the differences in spinal stability and stabilizing potential of instrumentation after cervical corpectomy and spondylectomy.

Methods: Seven human cadaveric specimens were tested: 1) intact; 2) after grafted C5 corpectomy and anterior C4-C6 plate; 3) after adding posterior C4-C6 screws/rods; 4) after extending posteriorly to C3-C7; 5) after grafted C5 spondylectomy, anterior C4-C6 plate, and posterior C4-C6 screws/rods; and 6) after extending posteriorly to C3-C7. Pure moments induced flexion, extension, lateral bending, and axial rotation; angular motion was recorded optically.

Results: After corpectomy, anterior plating alone reduced the angular range of motion to a mean of 30% of normal, whereas added posterior short- or long-segment hardware reduced range of motion significantly more (P < 0.003), to less than 5% of normal. Constructs with posterior rods spanning C3-C7 were stiffer than constructs with posterior rods spanning C4-C6 during flexion, extension, and lateral bending (P < 0.05), but not during axial rotation (P > 0.07). Combined anterior and C4-C6 posterior fixation exhibited greater stiffness after corpectomy than after spondylectomy during lateral bending (P = 0.019) and axial rotation (P = 0.001). Combined anterior and C3-C7 posterior fixation exhibited greater stiffness after corpectomy than after spondylectomy during extension (P = 0.030) and axial rotation (P = 0.0001).

Conclusion: Circumferential fixation provides more stability than anterior instrumentation alone after cervical corpectomy. After corpectomy or spondylectomy, long circumferential instrumentation provides better stability than short circumferential fixation except during axial rotation. Circumferential fixation more effectively prevents axial rotation after corpectomy than after spondylectomy.

Download full-text PDF

Source
http://dx.doi.org/10.1227/01.NEU.0000327569.03654.96DOI Listing

Publication Analysis

Top Keywords

axial rotation
24
corpectomy spondylectomy
16
anterior c4-c6
12
lateral bending
12
circumferential fixation
12
corpectomy
9
instrumentation cervical
8
cervical corpectomy
8
corpectomy anterior
8
c4-c6 plate
8

Similar Publications

Background: Hallux valgus (HV) is a complex, multiplanar deformity. In this study, we examined the interrelationships between various components of this deformity using weightbearing computed tomography (WBCT). We hypothesized that the severity of traditional axial plane deformities would correlate with malpositioning of the metatarsosesamoid complex, first-ray coronal rotational deformity, and malalignment of the hindfoot and midfoot.

View Article and Find Full Text PDF

Preserving Cervical Mobility: A Novel Robot-Assisted Approach for Atlas Fracture Fixation.

Am J Case Rep

January 2025

Department of Orthopedic Surgery, Sichuan Provincial People's Hospital, School of Medicine, University of Electronic Science and Technology of China, Chengdu, Sichuan, China.

BACKGROUND The management of unstable atlas fractures remains a subject of ongoing debate and controversy. The conservative surgical treatment commonly involves fusion, resulting in severe loss of cervical spine mobility, and a large incisions and extensive tissue dissection are required. We aim to introduce a novel concept and surgical approach for treating atlas fracture, one that involves minimizing trauma while maintaining mobility of the upper cervical spine without resorting to fusion.

View Article and Find Full Text PDF

Effect of Clamped Member Material and Thickness on Bolt Self-Loosening Under Transverse Loads.

Materials (Basel)

January 2025

Department of Mechanical Engineering, École de Technologie Supérieure, 1100 Notre-Dame Ouest, Montreal, QC H3C 1K3, Canada.

Bolted joints, prevalent in industrial applications for component fastening, are susceptible to self-loosening-a critical issue resulting in a gradual reduction in clamping force. Gaining insight into the underlying mechanisms of self-loosening is crucial. While prior research has largely focused on evaluating component stiffness, limited attention has been given to its impact on the self-loosening behavior of bolted joints under transverse cyclic loading.

View Article and Find Full Text PDF

Purpose: Currently, no gold standard exists for 3D analysis of virtually planned surgery accuracy postoperatively. The aim of this study was to present a new, validated and standardised methodology for 3D postoperative assessment of surgical accuracy in patients undergoing 3D virtually planned and guided corrective osteotomies.

Methods: All patients who underwent 3D planned corrective osteotomy in 2021-2022 at our center with a postoperative CT were included.

View Article and Find Full Text PDF

Fibrous dysplasia (FD) is a benign tumor condition in which normal bone is replaced by structurally deficient fibrous lamellar bone. It represents approximately 5-7% of benign bone tumors and occurs in two presentations: monostotic, which is the most common, and polyostotic. The proximal femur is one of the most common locations for benign tumors, including FD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!