Aim: Atrial undersensing (AUS) in single-lead VDD pacemakers may be due to diminished P-wave amplitude secondary to local inflammation beneath the electrodes closer to atrial wall. The aim of this study was to assess the potential effect of distance between atrial electrode and atrial wall on immediate and long-term atrial sensing stability in VDD systems.
Methods: A total of 275 patients with normal sinus node function who received VDD pacemakers were enrolled into the study and were followed up for a median duration of 33 months. During each control visit, a standard 12-lead electrocardiogram (ECG) was obtained and standard pacemaker function assessment was performed including testing for pacing threshold and atrioventricular synchrony. The distance between atrial electrode and atrial wall was measured from chest X-ray.
Results: Of the 275 patients, AUS was detected in 59 patients. Univariate predictors of AUS were use of closely spaced bipolar ring atrial electrode (CSBR) (P = 0.01), wider atrial ring-spacing (P = 0.03), and atrial sensitivity programmed to a higher level (P = 0.001). Use of CSBR (P = 0.04) and atrial sensitivity > or =0.3 mV (P = 0.02) were observed to be the independent predictors for AUS. When the distance between atrial electrode and atrial wall was <7 mm, AUS was less with diagonally arranged bipolar ring electrodes (DABR) than it was with CSBRs (P = 0.02).
Conclusions: The distance between atrial electrode and atrial wall does not appear to affect AUS incidence in VDD pacemakers. For VDD electrodes closer to atrial wall, AUS was significantly less likely in DABR-type electrodes.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6932008 | PMC |
http://dx.doi.org/10.1111/j.1542-474X.2008.00243.x | DOI Listing |
BMC Cardiovasc Disord
December 2024
Department of Cardiology, The First Affiliated Hospital of Dalian Medical University, Dalian, China.
Background: Delayed lead perforation is a rare complication of cardiac implantable electronic device (CIED). Clinical presentations range from completely asymptomatic to pericardial tamponade. Surgical lead extraction is recommended and transvenous lead extraction (TLE) with surgical backup is an alternative method.
View Article and Find Full Text PDFAnn Noninvasive Electrocardiol
January 2025
Department of Paediatrics and Adolescent Medicine, The University of Hong Kong, Pok Fu Lam, Hong Kong.
Background: Brugada syndrome (BrS) is an inherited channelopathy characterized by right precordial ST-segment elevation. This study investigates the clinical and genetic characteristics of children with BrS in Hong Kong.
Methods: A retrospective review was conducted at the only tertiary pediatric cardiology center in Hong Kong from 2002 to 2022, including all pediatric BrS patients under 18 years old.
JACC Clin Electrophysiol
November 2024
Department of Cardiology, Institute of Science Tokyo, Tokyo, Japan.
Background: Conventional endocardial mapping cannot fully elucidate Marshall bundle (MB)-related atrial tachycardia (AT).
Objectives: This study aimed to clarify the clinical and electrophysiological characteristics of MB-related AT definitively diagnosed using catheter insertion.
Methods: Forty-eight patients with AT who had previously undergone mitral isthmus ablation were enrolled in this study.
Ann Noninvasive Electrocardiol
January 2025
Cardiovascular Institute-Hospital Clínico San Carlos, Madrid, Spain.
We present the case of an 80-year-old female with acute pulmonary edema and a dual chamber pacemaker with intermittent short AV delays in the surface ECG after blocked premature atrial contractions (PACs). The behavior was consistent with the programmed Window of Atrial Rate Acceleration Detection (WARAD) and did not require further parameter modifications. As most cardiologists and emergency department physicians are not familiar with brand-specific algorithms, we believe that this case report will make these noncompetitive atrial pacing algorithms more accessible to non-cardiologists.
View Article and Find Full Text PDFNurs Stand
December 2024
Faculty of Science and Technology, Department of Natural Sciences, Middlesex University London, London, England.
Rationale And Key Points: A 12-lead electrocardiogram (ECG) is a painless procedure commonly undertaken in healthcare practice. An ECG machine uses cables or 'leads' attached with stick-on electrodes to specific parts of the body to create a series of ECG traces, which are then printed onto graph paper. This provides a recording of the cardiac electrical activity measured from different angles or perspectives.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!