AI Article Synopsis

Article Abstract

A novel on-line microdialysis sampling coupled with flame atomic absorption spectrometry (FAAS) with an attractive application is reported. Microdialysates perfused through implanted microdialysis probes were directly introduced into the flame atomizer of a FAAS system using 0.2% HNO(3) as carrier solution at a nebulizer uptake flow rate of 6mlmin(-1). The interval for each determination was 90s (60s sampling time, 10s read time and 20s washing time). The analytical characteristics of the on-line microdialysis-FAAS system were validated as follows: linearity range, 0-300mgl(-1); detection limit (3sigma, n=7), 0.53mgl(-1); precision (R.S.D., n=50), 4.1%. By comparing Mg levels in the blood of living rabbits with the results obtained from in vivo no net flux (NNF) method, the accuracy of the proposed on-line method was found to be good. The present method can be successfully applied to the in vivo monitoring of diffusible Mg in the blood of living rabbits after magnesium sulfate (MgSO(4)) administration with a temporal resolution of 1.5min.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.talanta.2004.12.022DOI Listing

Publication Analysis

Top Keywords

blood living
12
on-line microdialysis
8
microdialysis sampling
8
sampling coupled
8
coupled flame
8
flame atomic
8
atomic absorption
8
absorption spectrometry
8
vivo monitoring
8
monitoring diffusible
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!