Spectrophotometric and high performance liquid chromatographic (HPLC) methods have been developed for the determination of cis-platin and carboplatin based on the pre-column derivatization of platinum(II) with 2-acetylpyridine-4-phenyl-3-thiosemicarbazone. The complex was extracted in chloroform with molar absorptivity of 2.2x10(4)Lmol(-1)cm(-1) at 380nm. The complex eluted from a Phenomenex C-18 (150mmx4.6mm i.d.) column with methanol:water:acetonitrile:tetrabutyl ammonium bromide (1mM) (44:30:25:1, v/v/v/v) with a flow rate of 1ml/min and UV detection at 260nm. Ruthenium(IV) and selenium(IV) also separated completely. The linear calibration curve was with 0.5-12.5microg/ml and detection limit of 10ng/ml platinum(II).The analysis of cis-platin and carboplatin injections by spectrophotometric and HPLC methods indicated relative standard deviation (R.S.D.) of 0.66-2.1%. The method was used for the determinations of cis-platin in serum and urine of cancer patients after chemotherapy and platinum contents were found 148-444 and 50-90ng/ml with R.S.D. of 0.3-3.0 and 0.6-2.4% for the serum and urine, respectively. The recovery of platinum(II) from serum was 97% with R.S.D. 2.2%.