Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: 1) To determine the relationship between transcutaneous bilirubin measurements (TcB), performed using BiliCheck or Minolta Air-Shields JM-103, and total serum bilirubin levels (TSB) and 2) to evaluate the predictive accuracy of TcB measurements for risk using the nomogram recommended by the Canadian Pediatric Society (CPS).
Methods: A total of 154 healthy term neonates from the newborn nursery at McMaster Children's Hospital meeting the inclusion criteria were enrolled. TcB measurements were performed within 30 min of obtaining the TSB measurement.
Results: Both devices showed a good correlation with the laboratory method (BiliCheck-Vitros, R2=0.86; Minolta Air-Shields JM-103-Vitros, R2=0.85), but underestimated the serum bilirubin. Applying the risk classification using the 40th, 75th, and 95th percentile of the Bhutani nomogram a 6%, 0%, and 1% false negative rate was found for BiliCheck and 62%, 74% and 81% for the Minolta Air-Shields JM-103 device. After correcting for the differences using either the bias or the 95% CI the false negative rate was reduced to zero in all cases.
Conclusion: TcB measurements cannot be directly applied to a TSB nomogram but must be adjusted for any observed biases in order to avoid misclassifying newborns at risk for hyperbilirubinemia.