Design-of-experiment optimization of exhaled breath condensate analysis using a miniature differential mobility spectrometer (DMS).

Anal Chim Acta

Department of Mechanical and Aeronautical Engineering, University of California, Davis, One Shields Avenue, Davis, CA 95616, United States.

Published: November 2008

AI Article Synopsis

  • Analytical instruments are valuable for diagnosing diseases by measuring small amounts of chemicals in complex biological samples, like exhaled breath.
  • The study focuses on using gas chromatography-differential mobility spectrometry (GC/DMS) to analyze human exhaled breath condensate, aiming for non-invasive disease diagnostics.
  • A factorial design-of-experiments was used to optimize the sensor by testing various parameter combinations to maximize the detection of acetone, a clinical biomarker, while ensuring the sensitivity to other chemical compounds is maintained.

Article Abstract

Analytical instruments that can measure small amounts of chemicals in complicated biological samples are often useful as diagnostic tools. However, it can be challenging to optimize these sensors using actual clinical samples, given the heterogeneous background and composition of the test materials. Here we use gas chromatography-differential mobility spectrometry (GC/DMS) to analyze the chemical content of human exhaled breath condensate (EBC). Ultimately, this system can be used for non-invasive disease diagnostics. Many parameters can be adjusted within this instrument system, and we implemented a factorial design-of-experiments to systematically test several combinations of parameter settings while concurrently analyzing effects and interactions. We examined four parameters that affect sensitivity and detection for our instrument, requiring a 2(4) factorial design. We optimized sensor function using EBC samples spiked with acetone, a known clinical biomarker in breath. Two outputs were recorded for each experiment combination: number of chemicals detected, and the amplitude of acetone signal. Our goal is to find the best parameter combination that yields the highest acetone peak while also preserving the largest number of other chemical peaks in the spectra. By optimizing the system, we can conduct further clinical experiments with our sensor more efficiently and accurately.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2600869PMC
http://dx.doi.org/10.1016/j.aca.2008.09.010DOI Listing

Publication Analysis

Top Keywords

exhaled breath
8
breath condensate
8
design-of-experiment optimization
4
optimization exhaled
4
condensate analysis
4
analysis miniature
4
miniature differential
4
differential mobility
4
mobility spectrometer
4
spectrometer dms
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!