The new selective access analysis system BM/Hitachi 917 was evaluated in an international multicentre study, mainly according to the ECCLS protocol for the evaluation of analysers in clinical chemistry. Forty-three different analytes, covering 56 different methods--enzymes, substrates, electrolytes, specific proteins, drugs and urine applications--were tested in seven European clinical chemistry laboratories. Additionally, the practicability of the BM/ Hitachi 917 was tested according to a standardized questionnaire. Within-run CVs (median of 3 days) for enzymes, substrates and electrolytes were <2% except for creatine-kinase MB isoform and lipase at low concentration. For proteins, drugs and urine analytes the within-run CVs were < 4% except for digoxin and albumin in urine. Between-day median CVs were generally < 3% for enzymes, substrates and electrolytes, and < 6% for proteins, drugs and urine analytes, except for lipase, creatine kinase and MB isoform, D-dimer, glycosylated haemoglobin, rheumatoid factors, digoxin, digitoxin, theophylline and albumin in urine in some materials. Linearity was found according to the test specifications or better and there were no relevant effects seen in drift and carry-over testing. The interference results clearly show that also for the BM/Hitachi 917 interference exists sometimes, as could be expected because of the chemistries applied. It is a situation that can be found in equivalent analysers as well. The accuracy is acceptable regarding a 95-105% recovery in standard reference material, with the exception of the creatinine Jaffé method. Most of the 160 method comparisons showed acceptable agreement according to our criteria: enzymes, substrates, urine analytes deviation of slope +/- 5%, electrolytes +/- 3%, and proteins and drugs +/- 10%. The assessment of practicability for 14 groups of attributes resulted in a grading of one-three scores better for the BM/Hitachi 917 than the present laboratory situation. In conclusion, the results of the study showed good analytical performance and confirmed the usefulness of the system as a consolidated workstation in medium-sized to large clinical chemistry laboratories.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548263PMC
http://dx.doi.org/10.1155/S1463924600000080DOI Listing

Publication Analysis

Top Keywords

analysis system
8
clinical chemistry
8
substrates electrolytes
8
multicentre evaluation
4
evaluation boehringer
4
boehringer mannheim/hitachi
4
mannheim/hitachi 917
4
917 analysis
4
system selective
4
selective access
4

Similar Publications

Objectives: We examined the health-related quality of life (HRQoL) during menopause transition (MT) among middle-aged Korean women.

Methods: This cross-sectional study comprised 2,290 middle-aged women who completed web-based questionnaires between 2020 and 2022. Based on self-reported menstrual cycle patterns, menopause status was classified as premenopausal, early or late transition, or postmenopausal.

View Article and Find Full Text PDF

Background: Low back pain (LBP) is highly prevalent and disabling, especially in agriculture sectors. However, there is a gap in LBP prevention and intervention studies in these physically demanding occupations, and to date, no studies have focused on horticulture workers. Given the challenges of implementing interventions for those working in small businesses, self-management offers an attractive and feasible option to address work-related risk factors and manage LBP.

View Article and Find Full Text PDF

Background: Health authorities worldwide have invested in digital technologies to establish robust information exchange systems for improving the safety and efficiency of medication management. Nevertheless, inaccurate medication lists and information gaps are common, particularly during care transitions, leading to avoidable harm, inefficiencies, and increased costs. Besides fragmented health care processes, the inconsistent incorporation of patient-driven changes contributes to these problems.

View Article and Find Full Text PDF

Background: Laparoscopic surgery training is a demanding process requiring technical and nontechnical skills. Surgical training has evolved from traditional approaches to the use of immersive digital technologies such as virtual, augmented, and mixed reality. These technologies are now integral to laparoscopic surgery training.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!