Background And Purpose: Polyethylene is commonly employed for bearings in acetabular cups used in hip replacements. Assessment of in vivo wear is important for evaluation and monitoring of wear in individual patients, as well as in different implant designs. Polyethylene wear is quantified by comparisons of radiographic measurements made on sequential pelvic radiographs. Variations in pelvic orientation and variation in direction of wear may cause underestimation of polyethylene wear measurements. The purpose of this study was to quantify these effects on 2-dimensional measurements of polyethylene wear.
Methods: A computer program designed to simulate radiographs was employed to generate virtual radiographs of a virtual pelvis with a total hip replacement. Effects caused by variation in pelvic spatial orientation and variations in wear direction on wear measurements were analyzed separately. A Monte Carlo computational algorithm was employed to describe the combined effects of these two factors.
Results: Variation in pelvic orientation induced a mean underestimation of wear of 0.4% (0-2.6). Variation in direction of wear introduced a mean underestimation of 8.5% (0-42). A mean underestimation of wear of 9% (0-99) was found when varying both pelvic orientation and direction of wear simultaneously.
Interpretation: Errors caused by variations in pelvic orientation and wear direction are likely to be small compared to other sources of error when performing polyethylene wear measurements in acetabular components.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/17453670810016623 | DOI Listing |
J Funct Morphol Kinesiol
December 2024
Institut de Biomécanique Humaine Georges Charpak, Arts et Métiers Sciences and Technologies, 75013 Paris, France.
The handstand is an exercise performed in many sports, either for its own sake or as part of physical training. Unlike the upright bipedal standing posture, little is known about the sagittal alignment and balance of the spine during a handstand, which may hinder coaching and reduce the benefits of this exercise if not performed correctly. The purpose of this study was to quantify the sagittal alignment and balance of the spine during a handstand using radiographic images to characterize the strategies employed by the spino-pelvic complex during this posture.
View Article and Find Full Text PDFAnat Rec (Hoboken)
December 2024
Biology Department, Northland Pioneer College, Holbrook, Arizona, USA.
Nothronychus graffami was a large therizinosaur represented by a single well-preserved individual from the Turonian Tropic Shale of southern Utah. It is characterized by an enlarged abdomen, small tail, and an extensively pneumatized axial skeleton, and is frequently regarded as herbivorous. Given the overall tail reduction and the development of a wide fused synsacrum with widely spaced acetabulae, it is reconstructed with an anteriorly rotated femur and a displaced resting ground reaction force anterior to the center of mass.
View Article and Find Full Text PDFSICOT J
December 2024
The Hazeley Academy, Emperor Drive, Hazeley, Milton Keynes, MK8 0PT, United Kingdom.
Introduction: Robot-assisted instrumentation during total hip arthroplasty (THA) has the potential to improve acetabular cup positioning. This study aimed to evaluate the precision of robotic-arm-assisted THA (rTHA) and assess whether the system can provide accurate cup positioning comparable to conventional THA (cTHA).
Methods: A single-center prospective cohort study consisting of 151 patients who underwent THA (108 rTHA and 43 cTHA).
Cureus
November 2024
Orthopaedic Department, Thriasio General Hospital, Athens, GRC.
The quadrilateral surface fractures of the acetabulum are becoming more frequent due to the aging population. The optimal fixation of the quadrilateral surface through an appropriate surgical approach and a reliable and effective technique presents a challenge for orthopedic trauma surgeons. In this study, we present the results of 12 patients treated in our department who underwent internal fixation of the quadrilateral surface with two reconstruction plates in a vertical orientation for adequate fracture buttressing.
View Article and Find Full Text PDFArch Orthop Trauma Surg
December 2024
Department of Orthopedics, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, 1095 Jie Fang Avenue, Wuhan, 430030, People's Republic of China.
Purpose: The purposes of this study was to develop a novel therapy-oriented acetabular bone defects classification system based on Statistical shape models (SSMs) and evaluate the reliability and reproducibility of its application.
Methods: We retrospectively reviewed and annotated pelvic Computed tomography (CT) of patients who had undergone revision total hip arthroplasty at our institution and constructed a dataset to develop the novel method for quantitatively describing acetabular bone defects that is based on SSMs. Then, using this method, classification systems are suggestive for diagnosis and corresponding treatment suggestion.