Validation of spectral domain optical coherence tomographic Doppler shifts using an in vitro flow model.

Invest Ophthalmol Vis Sci

UPMC Eye Center, Eye and Ear Institute, Ophthalmology and Visual Science Research Center, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania 15213, USA.

Published: February 2009

Purpose: To validate velocity measurements produced by spectral domain optical coherence tomography (SD-OCT) in an in vitro laminar flow model.

Methods: A 30-mL syringe filled with skim milk was inserted into a syringe pump. Intravenous (i.v.) tubing connected the syringe within the pump to a glass capillary tube (internal diameter, 0.579 mm) shallowly embedded in agarose gel, then to a collection reservoir. SD-OCT imaging was performed with an anterior segment eye scanner and optics engine coupled with a 100-nm bandwidth broadband superluminescent diode. Scan density of 128 x 128 A-scans was spread over a 4 x 4 mm area, and each A-scan was 2 mm in length. Fifteen sequential stationary A-scans were obtained at each 128 x 128 position, and Doppler shifts were calculated from temporal changes in phase. The beam-to-flow vector Doppler angle was determined from three-dimensional scans.

Results: In all reflectance and Doppler images, a clear laminar flow pattern was observed, with v(max) appearing in the center of the flow column. Phase wrapping was observed at all measured flow velocities, and fringe washout progressively shattered reflectance and phase signals beyond the Nyquist limit. The observed percentages of the velocity profile at or below Nyquist frequency was highly correlated with the predicted percentages (R(2)=0.934; P=0.007).

Conclusions: SD-OCT provides objective Doppler measurements of laminar fluid flow in an in vitro flow system in a range up to the Nyquist limit.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2916766PMC
http://dx.doi.org/10.1167/iovs.08-2305DOI Listing

Publication Analysis

Top Keywords

spectral domain
8
domain optical
8
optical coherence
8
doppler shifts
8
vitro flow
8
laminar flow
8
syringe pump
8
128 128
8
nyquist limit
8
flow
7

Similar Publications

Robust and time-resolved estimation of cardiac sympathetic and parasympathetic indices.

R Soc Open Sci

January 2025

Sorbonne Université, Paris Brain Institute (ICM), CNRS UMR7225, INRIA Paris, INSERM U1127, Hôpital de la Pitié Salpêtrière, AP-HP, Paris 75013, France.

The time-resolved analysis of heart rate (HR) and heart rate variability (HRV) is crucial for the evaluation of the dynamic changes of autonomic activity under different clinical and behavioural conditions. Standard HRV analysis is performed in the frequency domain because the sympathetic activations tend to increase low-frequency HRV oscillations, while the parasympathetic ones increase high-frequency HRV oscillations. However, a strict separation of HRV into frequency bands may cause biased estimations, especially in the low-frequency range.

View Article and Find Full Text PDF

Spectrally encoded flow cytometry using few-mode fiber collection.

Biomed Opt Express

January 2025

Faculty of Biomedical Engineering, Technion-Israel Institute of Technology, Haifa 3200003, Israel.

In fiber-based confocal microscopy, using two separate fibers for illumination and collection enables the use of a few-mode fiber to achieve an effect similar to opening the pinhole in a conventional confocal microscope. In some Fourier-domain applications, however, or when a spectral measurement is involved, the coherent light detection would lead to noticeable spectral modulation artifacts that result from differential mode delay, an effect caused by the multimode propagation in the collection fiber. After eliminating these artifacts by using mode-dependent polarization control, we demonstrate effective spectrally encoded imaging with improved signal efficiency and lower speckle noise, and only a minor, negligible reduction in lateral and axial resolutions.

View Article and Find Full Text PDF

Complete 2π cycling of a phase around a phase singularity leads to a rapid phase variation in the nearby zones and forms a sharp local -vector peak. In this paper, the intensity distribution in the spatial domain is transformed into a -vector distribution in the wave vector domain, and we prove that the local -vector peak is generated at the point of minimum light field intensity. The local -vector peak is sharper when the minimum point is closer to the phase singularity.

View Article and Find Full Text PDF

A recent advancement in distributed sensing known as the time-expanded phase-sensitive optical time-domain reflectometry (TE Φ-OTDR) addresses the trade-off between spatial resolution and detection bandwidth, enabling centimeter-scale resolution alongside RF detection bandwidth in the order of MHz. To date, TE Φ-OTDR approaches extract the fiber response from the first Nyquist zone (NZ). In this Letter, we propose a post-processing strategy to enhance the SNR by spectrally averaging different NZs.

View Article and Find Full Text PDF

NEMS generated electromechanical frequency combs.

Microsyst Nanoeng

January 2025

Department of Systems Design Engineering, University of Waterloo, Waterloo, ON, N2L 3G1, Canada.

This paper presents a novel technique for low-power generation of frequency combs (FC) over a wide frequency range. It leverages modal interactions between electrical and mechanical resonators in electrostatic NEMS operating in air to provide a simple architecture for FC generators. A biased voltage signal drives the electrical resonator at resonance which is set to match an integer submultiple of twice the mechanical resonator's resonance.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!