Gaseous microemboli detection in a simulated pediatric CPB circuit using a novel ultrasound system.

ASAIO J

Department of Pediatrics, Pediatric Cardiac Research Laboratories, Penn State Milton S. Hershey Medical Center, Penn State College Medicine, Penn State Children's Hospital, Hershey, Pennsylvania 17033-0850, USA.

Published: November 2008

The objective of this study was to evaluate the effect of flow rate and perfusion mode on the delivery of gaseous microemboli in a simulated pediatric cardiopulmonary bypass (CPB) circuit with an open arterial filter purge line using a novel ultrasound detection system. The circuit was primed with 450 ml fresh, heparinized bovine blood plus 200 ml Lactated Ringer's solution (total volume 650 ml, corrected Hct 25%). After the injection of 5 ml air into the venous line, an Emboli Detection and Classification (EDAC) Quantifier (Luna Innovations, Inc., Roanoke, VA) was used to simultaneously record microemboli counts at postpump, postoxygenator, and postarterial filter sites. Trials were conducted at four different flow rates (500, 750, 1,000, 1,250 ml/min) and two perfusion modes (pulsatile, nonpulsatile). Microemboli counts uniformly increased with increasing pump flow rates. In all trials, the majority of gaseous microemboli detected in the simulated pediatric CPB circuit were <20 microm in diameter. At the lowest flow rate tested (500 ml/min), all microemboli (>10 microm) were cleared from the circuit by the oxygenator and arterial filter. Clearance efficiency was decreased at higher flow rates (750-1,250 ml/min). Over 98% of microemboli detected at the postoxygenator site were <40 microm in diameter. In general, pulsatile flow delivered more microemboli to the circuit at postpump and postoxygenator sites than nonpulsatile flow. The results of this study confirmed that entrained air from the venous line could be delivered to the systemic circulation (as represented by our pediatric pseudo patient) at flow rates from 750 to 1,250 ml/min, despite the presence of an arterial filter and open arterial filter purge line. All of the microemboli distal to the arterial filter were smaller than the conventional detectable level of 40 microm.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0b013e318186d32aDOI Listing

Publication Analysis

Top Keywords

gaseous microemboli
12
simulated pediatric
12
cpb circuit
12
flow rates
12
pediatric cpb
8
novel ultrasound
8
arterial filter
8
microemboli counts
8
microemboli detected
8
circuit
5

Similar Publications

Randomized Trial of Pulsatile and Nonpulsatile Flow in Cyanotic and Acyanotic Congenital Heart Surgery.

World J Pediatr Congenit Heart Surg

December 2024

Penn State Hershey Pediatric Cardiovascular Research Center, Penn State College of Medicine, Hershey, PA, USA.

Background: The study objective was to determine the impact of cardiopulmonary bypass perfusion modalities on cerebral hemodynamics and clinical outcomes in congenital cardiac surgery patients stratified by acyanotic versus cyanotic heart disease.

Methods: A total of 159 pediatric (age <18 years) cardiac surgery patients were prospectively randomized to pulsatile or nonpulsatile cardiopulmonary bypass and stratified by type of congenital heart disease: acyanotic versus cyanotic. Intraoperative cerebral gaseous microemboli counts and middle cerebral artery pulsatility index were assessed.

View Article and Find Full Text PDF
Article Synopsis
  • * Results showed that NO significantly reduced both the number and volume of microemboli compared to a control group, with a decrease from 55,478 to 1,197 microemboli over one hour (p=0.016).
  • * Additionally, the group receiving NO exhibited lower levels of neuron-specific enolase, indicating reduced brain injury, with concentrations dropping to 7.7 ng/ml, compared to 11.2 ng/ml in the control group (p=0.047).
View Article and Find Full Text PDF
Article Synopsis
  • ECMO oxygenators for children need to handle varying flow rates, as patients range from newborns to teens, with some requiring gas: blood flow ratios higher than the typical 2:1 for effective carbon dioxide removal.
  • An evaluation of A.L.ONE™ ECMO oxygenators showed that increasing the gas: blood flow ratio did not lead to significant rises in gas phase pressures or cause harmful gaseous microemboli (GME) production, whether at different gas flow rates or oxygen concentrations.
  • The study concludes that, in an experimental setting, higher GBFRs do not result in increased gas phase pressures or GME release, ensuring safety in pediatric ECMO applications.
View Article and Find Full Text PDF

Background: During cardiac surgery the use of a minimal extracorporeal circulation (MiECC) system may reduce the adverse effects for the patient. This is probably caused by reduced inflammation and hemodilution. For the use of a MiECC circuit, a venous bubble trap (VBT) is warranted for safety reasons.

View Article and Find Full Text PDF

Objective: During cardiopulmonary bypass (CPB), gaseous microemboli (GME) that originate from the extracorporeal circuit are released into the arterial blood stream of the patient. Gaseous microemboli may contribute to adverse outcome after cardiac surgery with CPB. Possibly, air may be collected in the right atrium during induction of anesthesia and released during CPB start.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!