Purpose: Report the results of using a permanently implantable dosimeter in radiation therapy: determine specific adverse events, degree of migration, and acquire dose measurements during treatment to determine difference between expected and measured dose.

Methods And Materials: The Dose Verification System is a wireless, permanently implantable metal-oxide semiconductor field-effect transistor dosimeter using a bidirectional antenna for power and data transfer. The study cohort includes 36 breast (33 patients received two devices) and 29 prostate (21 patients received two devices) cancer patients. A total of 1,783 and 1,749 daily dose measurements were obtained on breast and prostate patients, respectively. The measurements were compared with the planned expected dose. Biweekly computed tomography scans were obtained to evaluate migration and the National Cancer Institute's Common Toxicity Criteria, version 3, was used to evaluate adverse events.

Results: Only Grade I/II adverse events of pain and bleeding were noted. There were only four instances of dosimeter migration of >5 mm from known factors. A deviation of > or =7% in cumulative dose was noted in 7 of 36 (19%) for breast cancer patients. In prostate cancer patients, a > or =7% deviation was noted in 6 of 29 (21%) and 8 of 19 (42%) during initial and boost irradiation, respectively. The two patterns of dose deviation were random and systematic. Some causes for these differences could involve organ movement, patient movement, or treatment plan considerations.

Conclusions: The Dose Verification System was not associated with significant adverse events or migration. The dosimeter can measure dose in situ on a daily basis. The accuracy and utility of the dose verification system complements current image-guided radiation therapy and intensity-modulated radiation therapy techniques.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2590783PMC
http://dx.doi.org/10.1016/j.ijrobp.2008.05.058DOI Listing

Publication Analysis

Top Keywords

prostate patients
12
radiation therapy
12
adverse events
12
dose verification
12
verification system
12
cancer patients
12
dose
10
breast prostate
8
permanently implantable
8
dose measurements
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!