Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: To evaluate mechanical and hematological compatibility of a pediatric, temporary left heart support system in a lamb model as a less traumatic alternative to the widely used ECMO.
Methods: A small, pulsatile rotary blood pump (target flow 0.5l/m at 80 mm Hg pressure head at 120 pulses per min) was inserted in six lambs (15.1+/-1 kg) via a left thoracotomy, through a purse string in the arcus aortae. With fluoroscopy the tip (=inflow) of the catheter was positioned in the outflow tract of the left ventricle. The outflow part was positioned immediately above the aortic valve. Animals were extubated at the end of the procedure. Mechanical and hematological parameters were followed for 14 days.
Results: Five animals survived a 2-week follow-up. One animal died because of empyema on day 6. Flow maintained stable (0.8+/-0.2l/m) in all animals during the evaluation period. Free hemoglobin as a parameter of hemolysis and hematocrit remained also stable. Necropsy revealed minimal fibrous reaction on one aortic valve leaflet in one animal and small hematoma formation in three. All animals showed mild signs of endothelial damage on the aortic arch at the level of the motor housing. One animal showed signs of old kidney infarction suggesting possible embolization during placement.
Conclusion: This newly developed, catheter based, pediatric heart support system generates a stable flow for 14 days without compromising hematological stability and with acceptable tissue damage due to positioning of the catheter.