Insufficient cure under the condition of high irradiance and short irradiation time.

Dent Mater

Research and Development, Bisco Inc., Schaumburg, IL 60193, USA.

Published: March 2009

Objectives: To investigate if and why a plasma arc curing (PAC) light tends to undercure methacrylate-based resins or resin composites.

Methods: Model dimethacrylate resins, commercial dental adhesives, and commercial resin composites were cured using a PAC light and a halogen light with the similar radiant exposures but different combinations of irradiance and irradiation time. The degree of double bond conversion (DC) was measured with FTIR spectroscopy and analyzed as a function of radiant exposure.

Results: The PAC light produced a lower DC than the halogen light for the model resin with the lowest viscosity and for three of the four adhesives. With a high irradiance, the PAC light could cure three of the four composites as thoroughly as its halogen counterpart. When the irradiance was reduced, however, three composites yielded a lower DC.

Conclusions: Insufficient cure by PAC lights or any curing lights with very high irradiance is likely to happen when too short an irradiation time is used. It is because under higher irradiance, the lifetime of free radicals is shorter.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.dental.2008.07.007DOI Listing

Publication Analysis

Top Keywords

pac light
16
high irradiance
12
irradiation time
12
insufficient cure
8
short irradiation
8
halogen light
8
three composites
8
irradiance
6
light
6
pac
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!