A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Transient inhibition of motor function induced by the Cavitron ultrasonic surgical aspirator during brain mapping. | LitMetric

Transient inhibition of motor function induced by the Cavitron ultrasonic surgical aspirator during brain mapping.

Neurosurgery

Department of Neurosurgery, University of Milan, Fondazione IRCCS Ospedale Maggiore Policlinico, Mangiagalli e Regina Elena, Milan, Italy.

Published: July 2008

Objective: We report, for the first time, the occurrence of interference between a Cavitron ultrasonic surgical aspirator (CUSA) and intraoperative brain mapping performed by direct electrical stimulation (DES).

Methods: Intraoperative polygraphic recordings (electrocorticogram and electromyogram) were gathered from a 44-year-old patient harboring a recurrent Grade II oligoastrocytoma operated on with the aid of a CUSA and DES.

Results: Simultaneous use of CUSA and DES at the subcortical level in proximity to the corticospinal tract brought about the abolition of previously evident motor responses. This abolition was fully reversible after the CUSA was turned off. An analogous pattern of motor response inhibition was evident when the DES was applied cortically and the CUSA was used subcortically close to motor pathways. Interestingly, the authors had already observed a similar phenomenon in many patients when the CUSA was used for resection of lesions located within or in proximity to subcortical language pathways. In this setting, the CUSA induced transient speech disturbances that were confirmed afterwards by the DES. This interference with language and motor mapping might be interpreted as a transitory inhibition of axonal conduction.

Conclusion: The clinical significance of this interference is relevant when the CUSA and DES are used simultaneously for motor mapping because the CUSA can decrease the sensitivity of the brain mapping technique. Further studies will be required to determine the neurophysiological mechanism underlying this interference.

Download full-text PDF

Source
http://dx.doi.org/10.1227/01.NEU.0000335087.85470.18DOI Listing

Publication Analysis

Top Keywords

brain mapping
12
cusa
9
cavitron ultrasonic
8
ultrasonic surgical
8
surgical aspirator
8
cusa des
8
motor mapping
8
motor
6
mapping
5
transient inhibition
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!