Objective: We compared body composition estimates using an eight-electrode, segmental, multiple-frequency bioelectrical impedance analysis (segmental MF-BIA) and dual x-ray absorptiometry (DXA) in a group of healthy adults with a range of body mass indexes (BMIs).
Methods: Percentage of body fat (%BF), fat-free mass, and fat mass assessed by DXA and segmental MF-BIA in 132 healthy adults were classified by normal (N; 18.5-24.9 kg/m(2)), overweight (OW; 25-29.9 kg/m(2)), and obese (OB; 30-39.9 kg/m(2)) BMI.
Results: Compared with DXA, segmental MF-BIA overestimated %BF in the OB BMI group (3.4%; P < 0.0001). MF-BIA overestimated %BF among men (0.75%; P < 0.006) and women (0.87%; P < 0.006) and underestimated it in the N BMI group (-1.56%; P < 0.0001); %BF was not different between methods in the OW BMI group. Error in %BF determined by segmental MF-BIA and DXA increased as %BF increased (r = 0.42, P < 0.0001). Waist circumference was the only significant predictor of systematic error in %BF between MF-BIA and DXA (r = 0.60, P < 0.0001).
Conclusion: Eight-electrode, segmental MF-BIA is a valid method to estimate %BF in adults with BMI classified as N and OW, but not as OB. Estimation of trunk resistance with current segmental MF-BIA devices may explain the underestimation of %BF in the adults with OB BMI. Further examination of the effect of waist circumference and body fat distribution on the accuracy of BIA measurements is warranted.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.nut.2008.07.004 | DOI Listing |
Introduction: Bioelectrical impedance analysis (BIA) systems are gaining popularity for use in research and fitness assessments as the technology improves and becomes more affordable and easier to use. Multifrequency BIA (MF-BIA) may improve accuracy and precision using octopolar contacts for segmental analyses.
Purpose: Evaluate reliability, biological variability, and accuracy of component measures (total body water, mass, and composition) of commercially available MF-BIA system (InBody 770, Cerritos, California, USA).
Int J Environ Res Public Health
November 2022
Department of Research and Development, Starbia Meditek Co., Ltd., Taichung 40227, Taiwan.
We investigated differences in body composition measurements for the whole body and limb segments in elite male wrestlers between results of multi-frequency bioelectrical impedance analyses (MF) and dual energy X-ray absorptiometry (DXA). Sixty-six elite male wrestlers from Taiwan were recruited. Wrestlers' body fat percentage (PBF), whole body fat-free mass (FFM), whole body lean soft tissue mass (LSTM), and fat-free mass of arms, legs and trunk (FM, FFM, FFM) were measured by MF and DXA, and analyzed using Pearson correlation coefficient and Bland-Altman plot.
View Article and Find Full Text PDFSpinal Cord
March 2022
Faculty of Behavioural and Movement Sciences, Vrije Universiteit Amsterdam, Amsterdam Movement Sciences, Amsterdam, The Netherlands.
Study Design: Cross-sectional study.
Objectives: This study: (1) investigated the accuracy of bioelectrical impedance analysis (BIA) and skinfold thickness relative to dual-energy X-ray absorptiometry (DXA) in the assessment of body composition in people with spinal cord injury (SCI), and whether sex and lesion characteristics affect the accuracy, (2) developed new prediction equations to estimate fat free mass (FFM) and percentage fat mass (FM%) in a general SCI population using BIA and skinfolds outcomes.
Setting: University, the Netherlands.
J Cachexia Sarcopenia Muscle
August 2021
National Center for Geriatrics and Gerontology, Aichi, Japan.
Background: The Asian Working Group for Sarcopenia (AWGS) 2019 recommended the use of dual-energy X-ray absorptiometry (DXA) or bioelectrical impedance analysis (BIA) to assess appendicular lean mass (ALM). AWGS, European Working Group on Sarcopenia in Older People 2 (EWGSOP2), Foundation for the National Institutes of Health Sarcopenia Project (FNIH), and International Working Group on Sarcopenia (IWGS) reported different cutoff values for sarcopenia. We aimed to validate these cutoff values in a Japanese population using DXA and two different devices of segmental multi-frequency BIA (MF-BIA).
View Article and Find Full Text PDFInt J Exerc Sci
December 2020
Department of Kinesiology and Physical Education, Northern Illinois University, DeKalb, IL, USA.
The purpose of this study was to evaluate the validity of whole body percent fat (%BF) and segmental fat-free mass (FFM) using multi-frequency bioelectrical impedance analysis (MF-BIA) and dual-energy x-ray absorptiometry (DEXA) in college-aged adults. Sixty-two participants male ( = 32) and female ( 30) completed MF-BIA and DEXA measurements following established pre-test guidelines. %BF and segmental FFM (right arm, left arm, trunk, right leg, and left leg) were collected and analyzed.