Objective: Navigation systems enable neurosurgeons to guide operations with imaging data. Sensor-based neuronavigation uses an electromagnetic field and sensors to measure the positions of the patient's brain anatomy and the surgical instruments. The aim of this investigation was to determine the accuracy level of sensor-based tracking in a large patient collection.

Methods: This study covers 250 patients operated upon during a continuous 5.5-year period. The patients had a wide range of indications and surgical procedures. The operations were performed with a direct current (DC) pulsed sensor-based electromagnetic navigation system. Four kinds of errors were measured: the fiducial registration error (FRE), the target registration error (TRE), brain shift, and the position error (PE). These errors were calculated for five subgroups of indications: target determination and trajectory guidance, functional navigation, skull base and neurocranium, determination of resection volume, and transnasal and transsphenoidal access.

Results: The overall mean FRE was 1.66mm (+/-0.61mm). The overall mean TREs were 1.33mm (+/-0.51mm) centroid and 1.59mm (+/-0.57mm) lesional. The overall mean brain shift for applicable cases was 1.61mm (+/-1.14mm). The overall mean PE was 0.92mm (+/-0.54mm).

Conclusions: By and large, modern sensor-based neuronavigation operates within an acceptable and commonplace degree of error. However, the neurosurgeon must remain critical in cases of small lesions, and must exert caution not to introduce further interference from metal objects or electromagnetic devices.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clineuro.2008.06.017DOI Listing

Publication Analysis

Top Keywords

sensor-based neuronavigation
12
registration error
8
brain shift
8
sensor-based
5
neuronavigation evaluation
4
evaluation large
4
large continuous
4
continuous patient
4
patient population
4
population objective
4

Similar Publications

Transcranial current stimulation (tCS) techniques have been shown to induce cortical plasticity. As an important relay in the motor system, the cerebellum is an interesting target for plasticity induction using tCS, aiming to modulate its excitability and connectivity. However, until now it remains unclear, which is the most effective tCS method for inducing plasticity in the cerebellum.

View Article and Find Full Text PDF

Objective: Transcranial magnetic stimulation (TMS) is a noninvasive method for analyzing cortical function. To utilize TMS for presurgical functional diagnostics, the magnetic impulse must be precisely targeted by stereotactically positioning the coil. The aim of this study was to evaluate the usefulness of TMS for operation planning when combined with a sensor-based electromagnetic navigation system (nTMS).

View Article and Find Full Text PDF

Objective: Navigation systems enable neurosurgeons to guide operations with imaging data. Sensor-based neuronavigation uses an electromagnetic field and sensors to measure the positions of the patient's brain anatomy and the surgical instruments. The aim of this investigation was to determine the accuracy level of sensor-based tracking in a large patient collection.

View Article and Find Full Text PDF

Objective: Intraoperative speech mapping has evolved into the "gold standard" for neurosurgical removal of lesions near the language cortex. The integration of neuronavigation into a multimodal protocol can improve the reliability of this type of operation, but most systems require rigid fixation of the patient's head throughout the operation. This article describes and evaluates a new noninvasively attached sensor-based reference tool, which can replace rigid pin fixation of the patient's head during awake craniotomies.

View Article and Find Full Text PDF

Purpose: Although computer- and image-guided surgical procedures are an improvement of frame-guided stereotaxy, many navigation systems still require rigid fixation of the patient's head throughout the operation. This study describes the clinical application of a technical modification that enables cranial navigation with "free head mobility" using CT and MR images as well as the calculated 3-D reconstruction models.

Material And Methods: A sensor-based electromagnetic neuronavigation system was expanded to allow the localization and position monitoring of several sensors within an electromagnetic field.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!