A sensitive static headspace gas chromatography (sHS-GC) method is developed and validated for the determination of residual isobutylene oxide (IBO)-a genotoxic starting material in a drug substance for Phase I clinical trial studies. The experimental parameters, such as headspace vial pressure, headspace oven temperature, vial equilibration time, column flow rate, and GC oven temperature programs are optimized. Under the optimal conditions, the recovery is between 89.3% and 102.4% for spiked samples at three levels of IBO concentration in triplicates in sample preparations. Limits of quantitation (LOQ) and detection (LOD) of the standard solutions are 0.048 and 0.018 microg/mL, respectively. Linear range from 0.018 to 6 microg/mL is obtained with a correlation coefficient of 0.9999. The method is applied to determine residual IBO in drug substance samples from different batches.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/chromsci/46.7.637 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!