Objective: Endotracheal tube air leak pressures are used to predict postextubation upper airway compromise such as stridor, upper airway obstruction, or risk of reintubation. To determine whether the absence of an endotracheal tube air leak (air leak test >/=30 cm H2O) measured during the course of mechanical ventilation predicts extubation failure in infants and children.

Design: Prospective, blinded cohort.

Setting: Multidisciplinary pediatric intensive care unit of a university hospital.

Patients: Patients younger than or equal to 18 yrs and intubated >/=24 hrs.

Interventions: The pressure required to produce an audible endotracheal tube air leak was measured within 12 hrs of intubation and extubation. Unless prescribed by the medical care team, patients did not receive neuromuscular blocking agents during air leak test measurements.

Measurements And Main Results: The need for reintubation (i.e., extubation failure) was recorded during the 24-hr postextubation period. Seventy-four patients were enrolled resulting in 59 observed extubation trials. The extubation failure rate was 15.3% (9 of 59). Seven patients were treated for postextubation stridor. Extubation failure was associated with a longer median length of ventilation, 177 vs. 78 hrs, p = 0.03. Extubation success was associated with the use of postextubation noninvasive ventilation (p = 0.04). The air leak was absent for the duration of mechanical ventilation (i.e., >/=30 cm H2O at intubation and extubation) in ten patients. Absence of the air leak did not predict extubation failure (negative predictive value 27%, 95% confidence interval 6-60). The air leak test was >/=30 cm H2O before extubation in 47% (28 of 59) of patients yet 23 patients extubated successfully (negative predictive value 18%).

Conclusions: An endotracheal tube air leak pressure >/=30 cm H2O measured in the nonparalyzed patient before extubation or for the duration of mechanical ventilation was common and did not predict an increased risk for extubation failure. Pediatric patients who are clinically identified as candidates for an extubation trial but do not have an endotracheal tube air leak may successfully tolerate removal of the endotracheal tube.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2782931PMC
http://dx.doi.org/10.1097/PCC.0b013e3181849901DOI Listing

Publication Analysis

Top Keywords

air leak
44
endotracheal tube
28
tube air
24
extubation failure
24
leak test
16
>/=30 h2o
16
extubation
14
mechanical ventilation
12
air
11
leak
11

Similar Publications

In the past 20 years, sulfur hexafluoride (SF) has been considered a highly reliable tracer for assessing modern water (< 65 yrs old) in groundwater. However, modern-air contamination may introduce complications in interpreting data obtained using current sampling methods. A new airtight methodology isolates the sample from modern ambient air; thus, returning more reproducible and reliable results when compared to two traditional (air-sensitive and non-airtight) methods.

View Article and Find Full Text PDF

Background: Transcatheter closure of percutaneous paravalvular leak (PVL) is a technically challenging procedure, especially after surgical mechanical valve replacements (SMVR), as the risk of interference with the prosthetic valve discs and the complex interventional techniques required for mitral PVL closure. Our study was designed to review the results with transcatheter closure of PVL after SMVR.

Methods: From January 2018 through December 2023, a total of 64 patients with PVL after SMVR underwent transcatheter closure with the help of preoperative 3-dimensional printing model and simulator for image evaluation.

View Article and Find Full Text PDF

Physiological basis of non-invasive ventilation in the newborn.

Semin Perinatol

December 2024

Department of Women and Children's Health, School of Life Course Sciences, Faculty of Life Sciences and Medicine, King's College London, London, United Kingdom; Neonatal Intensive Care Unit, University of Patras, Patras, Greece. Electronic address:

Non-invasive ventilation (NIV) is a form of respiratory support provided primarily to preterm born infants in an effort to avoid any endotracheal intubation or as a weaning step following invasive ventilation. In the context of the respiratory distress syndrome of the newborn, NIV could target and partially reverse specific pathophysiological phenomena, by improving alveolar recruitment and establishing adequate functional residual capacity. It can also assist in minimizing lung injury by avoiding excessive pressure delivery, which can be harmful for the developing lung.

View Article and Find Full Text PDF

Ex vivo lung perfusion (EVLP) is used to evaluate donor lungs prior to lung transplantation. Development of pulmonary edema during EVLP is generally thought to represent inflammatory breakdown of the air-fluid barrier and these lungs are declined for transplant. We present the case of a donor lung that underwent stapled wedge resection during cold storage for air leak and the subsequent development of profound (∼650 mL) pulmonary edema around the staple line during EVLP.

View Article and Find Full Text PDF

Background: The real-world safety of robotic resections after neoadjuvant chemoimmunotherapy remains poorly defined in patients with non-small cell lung cancer. Due to reported increased operative challenges after neoadjuvant immunotherapy, we aim to describe our early institutional experience and outcomes after robotic resection in this clinical context.

Methods: We performed a retrospective chart review of patients with non-small cell lung cancer who underwent a robotic lobectomy, comparing patients from June 1, 2022, through October 31, 2023, who were treated with neoadjuvant chemoimmunotherapy consistent with the Checkmate-816 protocol and a control group with upfront resection.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!