Sleeve sphincter of Oddi (SO) manometry: a new method for characterizing the motility of the sphincter of Oddi.

J Hepatobiliary Pancreat Surg

Department of General and Digestive Surgery, Flinders University, Flinders Medical Centre, Bedford Park, SA 5042, Australia.

Published: December 2008

Background/purpose: Perfused multilumen sphincter of Oddi (SO) manometry is accepted as the gold standard for diagnosis of SO dysfunction. However, this technique is associated with a relatively high incidence of post-procedure acute pancreatitis. In addition, triple-lumen manometry recordings may be difficult to interpret, as movement may produce artifacts. We have refined the development of a sleeve sensor for human SO manometry. This assembly aims to overcome the above limitations. In this study the accuracy of sleeve SO manometry (SOM) has been evaluated and compared with standard triple-lumen perfused SOM.

Methods: Patients undergoing SO manometric studies consented to having both standard triple-lumen and sleeve SOM. A total of 32 paired studies were performed in 29 patients. Diagnosis was made only from standard triple-lumen SOM and the patient treated accordingly. For each study, SO basal pressure, contraction, amplitude, and frequency were recorded.

Results: There was no statistically significant difference in the recordings of SO basal pressure, contraction, amplitude, and frequency between the two techniques. A strong correlation was demonstrated between SO basal pressure determined with the two catheters. The accuracy of sleeve SOM is comparable to standard triple-lumen SOM, with less movement artifact. One patient developed mild post-manometric pancreatitis.

Conclusions: The sleeve catheter records SO pressures with comparable values to standard triple-lumen SOM. The sleeve assembly potentially can replace the use of the perfused triple-lumen catheter for the objective diagnosis of SO dysfunction.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00534-007-1262-1DOI Listing

Publication Analysis

Top Keywords

standard triple-lumen
20
sphincter oddi
12
triple-lumen som
12
basal pressure
12
oddi manometry
8
diagnosis dysfunction
8
accuracy sleeve
8
sleeve som
8
pressure contraction
8
contraction amplitude
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!