Successful treatment of rotary pump thrombus with the glycoprotein IIb/IIIa inhibitor tirofiban.

J Heart Lung Transplant

Advanced Heart Failure and Cardiac Transplant Service, Royal Perth Hospital, Perth, Western Australia, Australia.

Published: August 2008

Despite advances in blood pump technology, thrombus formation within left ventricular assist devices (LVADs) is a life-threatening complication with few therapeutic options. A 38-year-old woman who underwent rotary LVAD implantation as a bridge to cardiac transplant developed labile flows (4 to >10 liters), associated with power spikes (4 to 12 watts) and an increase in plasma free hemoglobin (0.86 g/liter), consistent with pump thrombus at Day 140 post-LVAD implantation, despite thromboprophylaxis with aspirin and warfarin. Within 12 hours of commencing an intravenous infusion of tirofiban at a rate of 0.1 mug/kg/min, there were signs of improvement of pump dysfunction, and complete resolution was evident at Day 4 with, stable flows, power consumption and normalization of plasma free hemoglobin. Tirofiban may be considered as an alternative thrombolytic treatment strategy in rotary pump thrombus to avoid the need for LVAD replacement.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.healun.2008.05.015DOI Listing

Publication Analysis

Top Keywords

pump thrombus
12
rotary pump
8
plasma free
8
free hemoglobin
8
pump
5
successful treatment
4
treatment rotary
4
thrombus
4
thrombus glycoprotein
4
glycoprotein iib/iiia
4

Similar Publications

Background: Hepatic artery infusion pump (HAIP) chemotherapy is a locoregional treatment for intrahepatic malignancies. HAIPs are surgically implanted, and the catheter tip is typically inserted into a ligated gastroduodenal artery stump. Potential complications at the catheter insertion site include dehiscence, pseudoaneurysm or extravasation, and adjacent hepatic arterial stenosis and thrombosis.

View Article and Find Full Text PDF

Progress of extracorporeal centrifugal pumps for mechanical circulatory supports.

J Artif Organs

January 2025

Department of Artificial Organs, National Cerebral and Cardiovascular Center, 6-1 Kishibe-Shinmachi, Suita, Osaka, 5648565, Japan.

This review traces the evolution of centrifugal blood pumps in mechanical circulatory support (MCS) systems. Initially met with concerns over blood damage and thrombus formation, centrifugal pumps have become crucial components in ventricular assist devices (VADs) and extracorporeal membrane oxygenation (ECMO) due to their simplified drive mechanisms and adaptability. This paper outlines three generations of centrifugal pump development: first-generation pumps with sealing components, second-generation pumps utilizing pivot bearings, and third-generation pumps employing contactless bearings.

View Article and Find Full Text PDF

Importance: The Aspirin and Hemocompatibility Events With a Left Ventricular Assist Device in Advanced Heart Failure (ARIES-HM3) study demonstrated that aspirin may be safely eliminated from the antithrombotic regimen after HeartMate 3 (HM3 [Abbott Cardiovascular]) left ventricular assist device (LVAD) implantation. This prespecified analysis explored whether conditions requiring aspirin (prior percutaneous coronary intervention [PCI], coronary artery bypass grafting [CABG], stroke, or peripheral vascular disease [PVD]) would influence outcomes differentially with aspirin avoidance.

Objective: To analyze aspirin avoidance on hemocompatibility-related adverse events (HRAEs) at 1 year after implant in patients with a history of CABG, PCI, stroke, or PVD.

View Article and Find Full Text PDF

Left ventricular (LV) thrombus is a serious complication of myocardial infarction (MI) that can lead to a fetal systemic embolism. Although coronary artery bypass graft surgery (CABG) after MI is widely performed, to our knowledge, there are no reports of LV thrombus in the early postoperative period. Here, we report a rare case of a 70-year-old man who underwent off-pump coronary artery bypass grafting (OPCAB) for unstable angina pectoris with reduced left ventricular ejection fraction (LVEF).

View Article and Find Full Text PDF

Background: Prior analyses have suggested that a smaller left ventricular end-diastolic diameter (LVEDD) is associated with reduced survival following HeartMate 3 left ventricular assist device implantation.

Objectives: In this trial-based comprehensive analysis, the authors sought to examine clinical characteristics and association with the outcome of this specific relationship.

Methods: The authors analyzed the presence of LVEDD <55 mm among 1,921 analyzable HeartMate 3 patients within the MOMENTUM 3 (Multicenter Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3) trial portfolio, on endpoints of overall survival and adverse events at 2 years.