Objectives: The aim of this study was to develop a simpler and more rapid analytical method for unbound teicoplanin in serum.

Methods: A new analytical method was developed by modifying an existing fluorescence polarization immunoassay (FPIA) method. The validation of the developed FPIA method was compared in the quantification of unbound teicoplanin with that of the high-performance liquid chromatography (HPLC) method reported previously. The developed FPIA method was employed for the measurement of 36 clinical samples collected from patients with methicillin-resistant Staphylococcus aureus (MRSA) infection.

Results: The limits of detection and quantification were 0.5 and 0.8 mug/mL, respectively. The recovery rate was 97.5-106.6%. The developed FPIA method showed better accuracy than the HPLC method. The within-run and interday reproducibility of the assay was good, with relative standard deviation values of 4.76-18.75% (within-run) and 5.68-13.95% (interday). Precision and accuracy of this method were within the acceptable limits defined in the US FDA Guidance for bioanalytical method validation. The correlation between the developed FPIA method and the HPLC method was good (r(2) = 0.87). A positive bias with the FPIA method was observed from the result of the Bland-Altman difference plot.

Conclusion: We firmly believe that the present method is useful for the adjustment of teicoplanin dosages for patients under various conditions.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1365-2710.2008.00923.xDOI Listing

Publication Analysis

Top Keywords

fpia method
24
developed fpia
16
method
14
unbound teicoplanin
12
hplc method
12
fluorescence polarization
8
polarization immunoassay
8
analytical method
8
method validation
8
fpia
6

Similar Publications

2,4-Dichlorophenoxyacetic acid (2,4-D) is one of the popular herbicides that is widely used in agriculture and can be found in food and water. A rapid and sensitive fluorescence polarization immunoassay (FPIA) was proposed for the detection of 2,4-D in juice and water. New tracers, 2,4-D-buthylenediamin fluoresceinthiocarbamyl (2,4-D-BDF) and 2,4-D-glycine aminofluorescein (2,4-D-GAF), were obtained and characterized.

View Article and Find Full Text PDF

Introduction: Healthcare systems face several challenges, with microbial infections being one of the main concerns. Therapeutic drug monitoring (TDM) is a strategy that has been encouraged to optimize antimicrobial regimens, particularly those with significant toxicity and narrow therapeutic indices, such as amikacin (AMK). We aimed to evaluate AMK concentrations of patients in a non-routine TDM setting and compare the performance of immunoassay and chromatography methods for routine clinical use.

View Article and Find Full Text PDF

This study explored the association between serum Hcy level and the all-cause mortality among osteoarthritis (OA) patients. This cohort study included patients diagnosed as OA from the National Health and Nutrition Examination Survey (NHANES) 1999-2006. Abbott Homocysteine assay, a fully automated fluorescence polarization immunoassay (FPIA) method, was used to measure the level of serum Hcy.

View Article and Find Full Text PDF

Fluorescence polarization (FP) assays are widely used to quantify biomolecules, and their combination with microfluidic devices has the potential for application in onsite analysis. However, the hydrophobic surface of polydimethylsiloxane (PDMS)-based microfluidic devices and the amphiphilicity of the blocking agents can cause the nonspecific adsorption of biomolecules, which in turn reduces the sensitivity of the FP assay. To address this, we demonstrated an FP assay with improved sensitivity in microfluidic devices using a polyethylene glycol-based surface modification to avoid the use of blocking agents.

View Article and Find Full Text PDF

Background: The American Academy of Orthopaedic Surgeons defined the acceptable threshold for elective safe surgery as a body mass index (BMI) under 40 due to the increased risk of complications. A consequence of this recommendation has been a hard cutoff based on BMI, which restricts access to care for an increasingly large and diverse population. There is an improved understanding that excess adipose tissue confers additional risk for postoperative complications, including infection, through mechanical and physiologic mechanisms.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!