Background: Despite the advances in total laboratory automation, a considerable amount of work in blood banks is still done using outdated manual methods. Some automated pre-transfusion testing instruments have recently been developed. Of these, we evaluated and compared the AutoVue Innova (Ortho, USA) and the Techno TwinStation (DiaMed AG, Switzerland).
Methods: Forward and reverse ABO/Rh typing and unexpected antibody screening and identification tests were performed on 4,628 samples using the manual method and the two automated instruments. Two different anticoagulants (EDTA and citrate) were compared in ABO/Rh typing and unexpected antibody screening tests. Titrating studies were conducted on the following 7 dilutions using 5 samples of irregular antibodies with anti-E, anti-E & -c, anti-D, and anti-Le(a) with anti-Fy(a): 1:2, 1:4, 1:8, 1:16, 1:32, 1:64, and 1:128. The test throughput per hour, the time required to perform 1 and 100 tests, and a simulation test for total events occurring in 1 day were also measured.
Results: No erroneous results were reported between the two instruments and the manual method. Discrepancies observed in 10 cases (0.4%) of ABO/Rh typing were of higher intensity with AutoVue Innova than with the manual method. AutoVue Innova exhibited the highest sensitivity in the titrating study and throughput performance compared with the manual method and the Techno TwinStation. Especially in the throughput and time required to complete 100 antibody screening tests, AutoVue Innova had a 3.3- and 3.5-fold higher performance, respectively, than Techno TwinStation.
Conclusions: Because both of the two fully automated instruments (AutoVue Innova and Techno TwinStation) had high levels of accuracy and performance, it is expected that use of fully automated instruments will reduce human labor, turnaround time, and operator error in the blood bank.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3343/kjlm.2008.28.3.214 | DOI Listing |
Indian J Hematol Blood Transfus
October 2019
Department of Laboratory Medicine, University of Ulsan College of Medicine, Ulsan University Hospital, Bangujin-sunhwandoro 877, Dong-gu, Ulsan, 44033 Korea.
The automated immunohematology analyzer IH-500 (Bio-rad, Cressier FR, Switzerland) was developed recently for blood bank tests and this study evaluated performance of IH-500. 200 blood samples for ABO/Rh typing were collected. ABO/Rh typing results measured by IH-500 was compared with conventional manual methods.
View Article and Find Full Text PDFTransfus Med
October 2017
Health Research Institute, Hospital Universitari I Politėcnic La Fe, Valencia, Spain.
Background: Implementation of fully automated analysers has become a crucial security step in the blood bank; it reduces human errors, allows standardisation and improves turnaround time (TAT).
Objectives: We aimed at evaluating the ease of use and the efficiency of the ORTHO VISION Analyser (VISION) in comparison to the ORTHO AutoVue Innova System (AutoVue) in six different laboratories.
Methods: After initial training and system configuration, VISION was used in parallel to AutoVue following the daily workload, both automates being based on ORTHO BioVue System column agglutination technology.
J Clin Diagn Res
March 2017
Senior Scientific Officer, Department of Transfusion Medicine, Max Superspeciality Hospital, Dehradun, Uttarakhand, India.
Introduction: The ultimate aim of pretransfusion testing is the acceptable survival of donor red cells in recipient's body and antibody detection plays a critical role in achieving the same. The cornerstone of antibody detection method is detecting an unexpected antibody as against the expected antibodies of ABO blood group system. Autoantibodies can also interfere with the detection of clinically significant alloantibodies.
View Article and Find Full Text PDFTransfus Apher Sci
August 2015
School of Biomedical sciences, Charles Stuirt University, Wagga Wagga, Australia. Electronic address:
This paper reports on an evaluation of the introduction of a blood bank automation system (Ortho AutoVue(®) Innova) in a hospital blood bank by considering the performance and workflow as compared with manual methods. The turnaround time was found to be 45% faster than the manual method. The concordance rate was found to be 100% for both ABO/Rh(D) typing and antibody screening in both of the systems and there was no significant difference in detection sensitivity for clinically significant antibodies.
View Article and Find Full Text PDFJ Blood Transfus
May 2014
Department of Laboratory Medicine, Central Hospital Steyr, Sierningerstraße 170, 4400 Steyr, Austria.
Unlabelled: The aim of this study was to determine the frequencies and specificities of "enzyme-only" detected red blood cell (RBC) alloantibodies in the routine antibody screening and antibody identification in patients hospitalized in Austria. Routine blood samples of 2420 patients were investigated. The antibody screening was performed with a 3-cell panel in the low-ionic strength saline- (LISS-) indirect antiglobulin test (IAT) and with an enzyme-pretreated (papain) 3-cell panel fully automated on the ORTHO AutoVue Innova System.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!