Stability measurements of 1-stage implants in the edentulous mandible by means of resonance frequency analysis.

Int J Oral Maxillofac Implants

Department of Prosthodontics, School of Dental Medicine, University of Bern, Bern, Switzerland.

Published: August 2008

AI Article Synopsis

Article Abstract

Objective: Resonance frequency analysis (RFA) is a method of measuring implant stability. However, little is known about RFA of implants with long loading periods. The objective of the present study was to determine standard implant stability quotients (ISQs) for clinical successfully osseointegrated 1-stage implants in the edentulous mandible.

Materials And Methods: Stability measurements by means of RFA were performed in regularly followed patients who had received 1- stage implants for overdenture support. The time interval between implant placement and measurement ranged from 1 year up to 10 years. The short-term group comprised patients who were followed up to 5 years, while the long-term group included patients with an observation time of > 5 years up to 10 years. For further comparison RFA measurements were performed in a matching group with unloaded implants at the end of the surgical procedure. For statistical analysis various parameters that might influence the ISQs of loaded implants were included, and a mixed-effects model applied (regression analysis, P <.0125).

Results: Ninety-four patients were available with a total of 205 loaded implants, and 16 patients with 36 implants immediately after the surgical procedure. The mean ISQ of all measured implants was 64.5 +/- 7.9 (range, 58 to 72). Statistical analysis did not reveal significant differences in the mean ISQ related to the observation time. The parameters with overall statistical significance were the diameter of the implants and changes in the attachment level. In the short-term group, the gender and the clinically measured attachment level had a significant effect. Implant diameter had a significant effect in the long-term group.

Conclusions: A mean ISQ of 64.5 +/- 7.9 was found to be representative for stable asymptomatic interforaminal implants measured by the RFA instrument at any given time point. No significant differences in ISQ values were found between implants with different postsurgical time intervals. Implant diameter appears to influence the ISQ of interforaminal implants.

Download full-text PDF

Source

Publication Analysis

Top Keywords

stability measurements
8
1-stage implants
8
implants edentulous
8
resonance frequency
8
frequency analysis
8
implant stability
8
implants
6
stability
4
measurements 1-stage
4
edentulous mandible
4

Similar Publications

How SNARE proteins generate force to fuse membranes.

Biophys J

January 2025

Department of Chemical Engineering, Columbia University, New York, NY 10027. Electronic address:

Membrane fusion is central to fundamental cellular processes such as exocytosis, when an intracellular machinery fuses membrane-enclosed vesicles to the plasma membrane for contents release. The core machinery components are the SNARE proteins. SNARE complexation pulls the membranes together, but the fusion mechanism remains unclear.

View Article and Find Full Text PDF

Massive endobronchial hemorrhage leading to Cardiac arrest during EBUS-TBNA: a case of successful resuscitation.

BMC Pulm Med

January 2025

Department of Pulmonary and Critical Care Medicine, School of Medicine, Zhongshan Hospital of Xiamen University, Xiamen University, Xiamen, Fujian, China.

Introduction: Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is commonly used for diagnosing mediastinal lymphadenopathy. Despite a low complication rate, severe hemorrhage can occur which is reported in this literature, particularly in hypervascular conditions like Castleman disease.

Methods: A 54-year-old male with idiopathic multicentric Castleman disease underwent EBUS-TBNA for mediastinal lymph node sampling.

View Article and Find Full Text PDF

Slopes influenced by multiple faults are prone to large-scale landslides triggered by multi-regional failures. Understanding the failure process and sequence is essential for the sustainable development of mining operations. This paper presents a method combining InSAR monitoring and numerical simulation to analyze the failure processes of slopes affected by multiple faults.

View Article and Find Full Text PDF

People with amblyopia show deficits in global motion perception, especially at slow speeds. These observers are also known to have unstable fixation when viewing stationary fixation targets, relative to healthy controls. It is possible that poor fixation stability during motion viewing interferes with the fidelity of the input to motion-sensitive neurons in visual cortex.

View Article and Find Full Text PDF

Objectives: Critically ill newborn infants often require simultaneous administration of multiple intravenous (IV) solutions through the same catheter lumen, making compatibility of these solutions crucial in neonatal intensive care units (NICUs). This study aimed to investigate the physical compatibility of insulin aspart, lidocaine, alprostadil and vancomycin with individualised two-in-one parenteral nutrition (PN).

Methods: The study was conducted at the hospital pharmacy's drug compounding facility of the University Medical Centre Utrecht.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!