[Relationship between measurements of tidal volume and thoracic inflating volume by pressure plethysmography in mice].

Zhejiang Da Xue Xue Bao Yi Xue Ban

Department of Pulmonology, The Second Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou 310009, China.

Published: May 2008

Objective: To evaluate the accordance of tidal volume (TV) with thoracic inflating volume (TIV) measured by pressure whole-body or double-chamber plethysmography in mice.

Methods: TV and TIV by double-chamber plethysmography were simultaneously measured in 6 mice with spontaneous respiratory frequencies 90 - 120/min; TV and TIV by whole-body plethysmography were measured in 8 paralyzed mice ventilated by fixed frequency.

Result: TIV value by double-chamber plethysmography was significantly higher than TV [(0.369 +/- 0.014) ml vs (0.356 +/- 0.012) ml, P < 0.01)] in 6 spontaneously breathing mice. TIV value by whole-body plethysmography was significantly lower than TV[(0.233 +/- 0.003) ml vs 0.3 ml, P < 0.001] in 8 paralyzed mice.

Conclusion: The TV value assessed by TIV measurement can not be accurate, because of the humidifying and heating of inhaled air and negative thoracic pressure during the measurement.

Download full-text PDF

Source
http://dx.doi.org/10.3785/j.issn.1008-9292.2008.03.007DOI Listing

Publication Analysis

Top Keywords

double-chamber plethysmography
12
tidal volume
8
volume thoracic
8
thoracic inflating
8
inflating volume
8
tiv double-chamber
8
tiv whole-body
8
whole-body plethysmography
8
plethysmography
6
tiv
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!