In a previous study on peel adhesion for medical tapes, it was shown that a stainless steel (SS) substrate better discriminated among medical tapes than a high-density polyethylene (HDPE) substrate. The objective of this study was to determine if a SS substrate would also better distinguish among transdermal drug delivery systems (TDDSs). Five TDDSs (Vivelle Dot, Climara, Catapres-TTS, Duragesic, and Mylan Fentanyl) were evaluated on three different substrates (SS, HDPE, and human cadaver skin). All measurements were made using a dwell time of approximately 3 min, a peel angle of 90 degrees, and a peel speed of 300 mm/min. Differences among TDDSs were greater for SS than for HDPE, using the F statistic for testing for differences among TDDSs means as a measure of heterogeneity, thereby indicating greater discrimination by SS.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jbm.b.31149 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!