Objective: Patient-ventilator synchrony during non-invasive pressure support ventilation with the helmet device is often compromised when conventional pneumatic triggering and cycling-off were used. A possible solution to this shortcoming is to replace the pneumatic triggering with neural triggering and cycling-off-using the diaphragm electrical activity (EA(di)). This signal is insensitive to leaks and to the compliance of the ventilator circuit.

Design: Randomized, single-blinded, experimental study.

Setting: University Hospital. PARTICIPANTS AND SUBJECTS: Seven healthy human volunteers.

Interventions: Pneumatic triggering and cycling-off were compared to neural triggering and cycling-off during NIV delivered with the helmet.

Measurements And Results: Triggering and cycling-off delays, wasted efforts, and breathing comfort were determined during restricted breathing efforts (<20% of voluntary maximum EA(di)) with various combinations of pressure support (PSV) (5, 10, 20 cm H(2)O) and respiratory rates (10, 20, 30 breath/min). During pneumatic triggering and cycling-off, the subject-ventilator synchrony was progressively more impaired with increasing respiratory rate and levels of PSV (p < 0.001). During neural triggering and cycling-off, effect of increasing respiratory rate and levels of PSV on subject-ventilator synchrony was minimal. Breathing comfort was higher during neural triggering than during pneumatic triggering (p < 0.001).

Conclusions: The present study demonstrates in healthy subjects that subject-ventilator synchrony, trigger effort, and breathing comfort with a helmet interface are considerably less impaired during increasing levels of PSV and respiratory rates with neural triggering and cycling-off, compared to conventional pneumatic triggering and cycling-off.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2517084PMC
http://dx.doi.org/10.1007/s00134-008-1163-zDOI Listing

Publication Analysis

Top Keywords

triggering cycling-off
16
pneumatic triggering
12
neural triggering
8
triggering
6
subject-ventilator synchrony
4
synchrony neural
4
neural versus
4
versus pneumatically
4
pneumatically triggered
4
triggered non-invasive
4

Similar Publications

Waveforms-guided cycling-off during pressure support ventilation improves both inspiratory and expiratory patient-ventilator synchronisation.

Anaesth Crit Care Pain Med

December 2022

Anaesthesia and Intensive Care, Emergency Department, Fondazione IRCCS Policlinico S. Matteo, Viale Golgi 19, 27100, Pavia, Italy.

Objective: To test the performance of a software able to control mechanical ventilator cycling-off by means of automatic, real-time analysis of ventilator waveforms during pressure support ventilation.

Design: Prospective randomised crossover study.

Setting: University Intensive Care Unit.

View Article and Find Full Text PDF

Automatic Adjustment of the Inspiratory Trigger and Cycling-Off Criteria Improved Patient-Ventilator Asynchrony During Pressure Support Ventilation.

Front Med (Lausanne)

November 2021

Jiangsu Provincial Key Laboratory of Critical Care Medicine, Department of Critical Care Medicine, Zhongda Hospital, School of Medicine, Southeast University, Nanjing, China.

Patient-ventilator asynchrony is common during pressure support ventilation (PSV) because of the constant cycling-off criteria and variation of respiratory system mechanical properties in individual patients. Automatic adjustment of inspiratory triggers and cycling-off criteria based on waveforms might be a useful tool to improve patient-ventilator asynchrony during PSV. Twenty-four patients were enrolled and were ventilated using PSV with different cycling-off criteria of 10% (PS), 30% (PS), 50% (PS), and automatic adjustment PSV (PS).

View Article and Find Full Text PDF

Comparative bench study evaluation of a modified snorkeling mask used during COVID-19 pandemic and standard interfaces for non-invasive ventilation.

Pulmonology

January 2023

Department of Emergency, Intensive Care Medicine and Anesthesia, Fondazione Policlinico Universitario A. Gemelli IRCCS, Rome, Italy; Istituto di Anestesiologia e Rianimazione, Università Cattolica del Sacro Cuore, Rome, Italy.

Purpose: The aim of this bench study is to compare the standard NIV and nCPAP devices (Helmet, H; Full face mask, FFM) with a modified full face snorkeling mask used during COVID-19 pandemic.

Methods: A mannequin was connected to an active lung simulator. The inspiratory and expiratory variations in airways pressure observed with a high simulated effort, were determined relative to the preset CPAP level.

View Article and Find Full Text PDF

Background: Pendelluft phenomenon is defined as the displacement of gas from a more recruited nondependent (ND) lung region to a less recruited dependent (D) lung region. This phenomenon may cause lung injury. Thus, a lung model for pendelluft was established, and the effects of ventilatory settings on pendelluft were examined.

View Article and Find Full Text PDF

Ventilator- and interface-related factors influencing patient-ventilator asynchrony during noninvasive ventilation.

Ann Thorac Med

January 2020

Department of Medicine, The University Sleep Disorders Center, College of Medicine, King Saud University, Riyadh, Saudi Arabia.

Patient-ventilator asynchrony (PVA) is common in patients receiving noninvasive ventilation (NIV). This occurs primarily when the triggering and cycling-off of ventilatory assistance are not synchronized with the patient's inspiratory efforts and could result in increased work of breathing and niv failure. In general, five types of asynchrony can occur during NIV: ineffective inspiratory efforts, double-triggering, auto-triggering, short-ventilatory cycling, and long-ventilatory cycling.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!