A comparative contrast-detail study of five medical displays.

Med Phys

Duke Advanced Imaging Laboratories, Department of Radiology, Duke University, Durham, North Carolina 27710, USA.

Published: April 2008

The objective of this study was to compare the contrast-detail performance of five different commercial liquid crystal displays (LCDs) to other LCD and cathode-ray tube (CRT) displays for medical applications. A contrast-detail phantom, supplemented with 5 in. of acrylic, was imaged on a commercial digital radiographic system using techniques comparable to chest radiography. The phantom design enabled observer evaluation by a four-alternative forced choice paradigm. The acquired images were independently scored by five observers on five medical display devices: a 5 megapixel monochrome LCD, a 3 megapixel monochrome LCD, a 9 megapixel color LCD, a 5 megapixel monochrome CRT, and a mammographic-grade monochrome CRT. The data were analyzed using the method suggested by the manufacturer based on a nearest neighbor correction technique. They were further analyzed using a logistic regression response model with a natural threshold using an overall chi-square test for display type followed by pairwise comparisons for individual display performance. The differences between the display devices were small. The standard analysis of the results based on the manufacturer-recommended method did not yield any statistically discernible trend among displays. The logistic regression analysis, however, indicated that the 5 megapixel monochrome LCD was statistically significantly (p <0.0001) superior to the others, followed by the 3 megapixel monochrome LCD (p<0.0001). The three other displays exhibited lower but generally similar performance characteristics. The findings suggest that 5 and 3 megapixel monochrome LCDs provide comparable but subtly superior contrast detectability than other tested displays, with the former performing slightly better in the detection of subtle and fine details.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.2868780DOI Listing

Publication Analysis

Top Keywords

megapixel monochrome
16
monochrome lcd
12
lcd megapixel
12
display devices
8
monochrome crt
8
logistic regression
8
lcd
5
megapixel
5
monochrome
5
comparative contrast-detail
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!