Objective: Gastric inflation (GI) is a significant issue when ventilation is performed on unprotected airways.
Design: Experimental analysis on the respiratory effects of hose extended bag-valve ventilation devices designed to reduce inspiratory pressure and flow.
Setting: Laboratory with lung/oesophageal sphincter simulator and pressure-flow-volume analyser. Lung compliance: 300ml/kPa, airway resistance: 0.5kPa/l/s. Lower oesophageal sphincter pressure (LOSP): 0.5kPa.
Interventions: Bag-valve ventilation of lung simulator. Twelve academic dental staff members used four devices: Ambu Mark III attached to either a reservoir bag (R) or a pressure relief valve (SV), SMART BAG (SB), and Easy Grip (EG) as control.
Results: After Bonferroni correction (p-level of significance 0.0083) for multiple comparisons, no evidence of difference between inspiratory tidal volumes (TVIN) administered by use of R (median 137ml) and SB (149ml) was found. Differences in TVIN were only detected between R and SV (188ml) (p=0.002). Only a trend towards TVIN differences between SB and R in comparison to EG (195ml) was found (p=0.009). Distributions of peak pressures differed when R (median 0.7kPa) and SV (1.0kPa) (p=0.006) or SB (0.7kPa) and SV (p=0.002) were compared. Peak inspiratory flow rates differed between EG (median 59l/min) and R (32l/min) as well as SB (42l/min) and between SB and SV (50l/min) (all with p=0.001). GI was lowest by use of R (median 103ml) compared to all other devices (EG: 518ml, SV: 394ml, SB: 271ml) (p=0.001). The areas under the pressure/flow over time curves were larger during SB compared to R ventilation. Mean airway pressures were significantly lower by use of R (0.1kPa) compared to SB (0.3kPa) (p<0.008).
Conclusion: Lowering GI by pressure-flow reduction may result in lower TV depending on the device used. Lowest GI resulted from R ventilation. This may be explained by the specific pressure/time or flow/time patterns achieved by use of this device.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.resuscitation.2008.02.022 | DOI Listing |
Sci Rep
January 2025
Emergency Department, The State Key Laboratory for Complex, Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.
Variation in the incidence, survival rate and factors associated with survival after cardiac arrest in China is reported. Some studies have tried to fill the knowledge gap regarding the epidemiology of cardiac arrest in China but were unable to identify reasons for the reported differences. Therefore, the purpose of this study was to describe Chinese management of cardiac arrest, particularly from the perspective of compression, ventilation, monitoring, treatment, and extracorporeal cardiopulmonary resuscitation.
View Article and Find Full Text PDFAnaesth Rep
January 2025
Department of Anaesthesia Rabin Medical Centre, Beilinson Hospital Petah Tikva Israel.
Venous thromboembolic disease remains a leading cause of maternal morbidity and mortality. We report a case of a 30-year-old woman at 37 gestation with a history of thalassaemia intermedia and splenectomy. During pregnancy, she had been managed with frequent blood transfusions and enoxaparin.
View Article and Find Full Text PDFJ Cardiothorac Vasc Anesth
December 2024
Department of Anesthesiology, Thomas Jefferson University, Philadelphia, PA.
Objectives: Following cardiac surgery, patients often require ventilatory support during transport to the intensive care unit (ICU). Manual ventilation using a bag valve mask (BVM) is commonly employed; however, mechanical ventilation may sometimes be preferred due to concerns regarding oxygenation, ventilation, and hemodynamic stability. The decision between manual and mechanical ventilation is typically based on clinical experience and surgical factors, as there is no established consensus or robust clinical evidence to guide this choice.
View Article and Find Full Text PDFClin Pract Cases Emerg Med
November 2024
Christiana Care Health System, Department of Emergency Medicine, Newark, Delaware.
Introduction: Propofol is an anesthetic agent commonly used in emergency department (ED) procedural sedation. It is often preferred in orthopedic procedures because of its muscle-relaxing properties. Rarely, however, it can induce agitation and muscle hypertonicity.
View Article and Find Full Text PDFWilderness Environ Med
December 2024
Department of Anesthesiology and Critical Care Medicine, Medical University of Innsbruck, Innsbruck, Austria.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!